MAN.TTB.EN--03--SDT270-Tanktest-User-Manual-Eng.Docx
15/17
12
Network settings
When your SDT270 is not connected to a Network, it is
advised to parameter the LAN to disable.This setting
permits to increase the battery autonomy of more than
20%.
Use the Right Arrow to highlight the LAN enabled field,
then using the Up Arrow set the field to NO.
Set the field to Yes when you want to connect your
SDT270 to a network via the Lemo15 socket.
Miscellaneous
13
Recommended calibration intervals
SDT recommends annual recalibration
of SDT instruments and annual verification of sensors.
This periodicity is coherent with the observed long-term stability of SDT electronic equipment’s.
However
-
For new equipment leaving the factory the initial period is extended to 15 months
instead of 12 (to avoid any possible discrimination between users due to transportation time,
customs clearance, etc.)
-
SDT recognizes that each customer may have their own Quality Assurance Standards and
internal requirements. If a customer has established a Quality Assurance Program that includes
a documented procedure for measurement deviations,
and if an instrument and the relevant sensor show acceptable deviations
and do not show signs of physical damage, then calibration periodicity could be extended to two
years.
-
Third party organizations having more authority (Class, or State Administrations) may
have defined their own regulations for example 6 months, or 2 years calibration interval:
by evidence, local legally applicable regulations supersede SDT rules.
For these reasons, SDT Calibration Certificates do no more mention a “calibration due date”,
but simply the “last calibration date”.
For user convenience, the date of the last calibration of an SDT200 or 270 instruments
can also be found in the menu screens.
This text refers to SDT document DC.QUAL.005
1
.
1
This document is part of SDT’s quality management system (ISO-9001 certified).