MAN.TTB.EN--03--SDT270-Tanktest-User-Manual-Eng.Docx  

15/17 

12

 

Network settings 

 

When your SDT270 is not connected to a Network, it is 
advised to parameter the LAN to disable.This setting 
permits to increase the battery autonomy of more than 
20%. 

Use the Right Arrow to highlight the LAN enabled field, 
then using the Up Arrow set the field to NO. 

Set the field to Yes when you want to connect your 
SDT270 to a network via the Lemo15 socket. 

 

Miscellaneous 

13

 

Recommended calibration intervals 

SDT recommends annual recalibration

 

of SDT instruments and annual verification of sensors. 

This periodicity is coherent with the observed long-term stability of SDT electronic equipment’s. 

However 

-

 

For new equipment leaving the factory the initial period is extended to 15 months 

instead of 12  (to avoid any possible discrimination between users due to transportation time, 

customs clearance, etc.)   

-

 

SDT recognizes that each customer may have their own Quality Assurance Standards and 

internal requirements.  If a customer has established a Quality Assurance Program that includes 

a documented procedure for measurement deviations,  

and if an instrument and the relevant sensor show acceptable deviations  

and do not show signs of physical damage, then calibration periodicity could be extended to two 

years. 

-

 

Third party organizations having more authority (Class, or State Administrations) may 

have defined their own regulations for example 6 months, or 2 years calibration interval:  

by evidence, local legally applicable regulations supersede SDT rules. 

For these reasons, SDT Calibration Certificates do no more mention a “calibration due date”,  

but simply the “last calibration date”. 

For user convenience, the date of the last calibration of an SDT200 or 270 instruments 
can also be found in the menu screens. 

This text refers to SDT document DC.QUAL.005

1

 

 

                                                            

1

 This document is part of SDT’s quality management system (ISO-9001 certified). 

Summary of Contents for SDT270 TankTest

Page 1: ...l Humanit 415 B 1190 Brussels Belgium Tel 32 0 2 332 32 25 email hearmore sdtultrasound com SDT North America 1532 Ontario Street Cobourg ON Phone 1 800 667 5325 1 905 377 1313 email info sdtultrasou...

Page 2: ...d To check your SDT270 version please refer to the chapter System Info Your SDT270 device proposes both text and icon interfaces This manual describes only the icon interface To switch between icon an...

Page 3: ...MAN TTB EN 03 SDT270 Tanktest User Manual Eng Docx 3 17 Table of Contents...

Page 4: ...wer source is its internal battery Using the charger is strictly forbidden Battery charge must exclusively be performed outside potentially explosive environments Do not use USB connectivity in danger...

Page 5: ...es these days leaving the 270 charging after the charger says it is charged does not hurt the battery in fact it very slowly tops the charge up a little more You can charge the SDT270 switched off You...

Page 6: ...ack TankTest sensor emerged sensor It corresponds to the TTB2 choice on the display The second one is equipped with a red collar This connector is meant for the red TankTest sensor immersed sensor It...

Page 7: ...urements M button to start the measurement acquisition Up and down arrows to increase and decrease amplification Left and right arrows to decrease and increase the audio volume of the headphones Enter...

Page 8: ...easured value is updated four times per second 6 1 1 Selecting a sensor Select a sensor by pressing the F1 button 6 1 2 Adjust the amplification Set up amplification using Up and Down arrow buttons ti...

Page 9: ...or has to take measurements separately for both red immersed and black emerged sensors When the operator decides to store measurement values when the system is under vacuum he should necessarily store...

Page 10: ...nd audio volume press F2 Using the arrow buttons highlight the audio recording mode icon and press the enter button 6 3 1 Set the record duration Press the F3 button Press the left and right arrow but...

Page 11: ...button Then you can select a measurement Ref or Test or an audio file Wave taking using the black sensor TTB1 or TTB2 or the red sensor TTR1 or TTR2 Then press the Enter button to display the list of...

Page 12: ...CD icon adjust the screen contrast and the operating time of the backlight Auto power down icon adjust the time before device auto power down when it is unused Reset icon force a factory reset of the...

Page 13: ...between icon and text interfaces using the up or down arrow keys 10 Keyboard functions when navigating inside the menus F1 to come back to the previous menu Left and right arrows to select the previo...

Page 14: ...the Down Arrow button to access to the next screen which indicates The last Calibration date The recommended next Calibration date The remaining days before the end of valid Calibration Press the Dow...

Page 15: ...n time customs clearance etc SDT recognizes that each customer may have their own Quality Assurance Standards and internal requirements If a customer has established a Quality Assurance Program that i...

Page 16: ...nty SDT International guarantees the SDT270 unit against manufacturing faults for a period of 2 two years with the exception of the battery and accessories charger headphones sensors etc these are gua...

Page 17: ...mitted to reproduce or duplicate in any form the whole or part of this document without the written permission of SDT International n v s a The information herein is believed to be accurate to the bes...

Reviews: