background image

C. CALIBRATION 

Before using the scale, please check the accuracy of the scale . If not 

accurate, please calibrate the scale with standard calibration weights.   

1 Turn on the scale for preheating 10 minutes. 

2  Make  sure  the  scale  is  unloaded  and  in  zero  mode,  long  press 

[TAR/CAL]  to  initiate  calibration  (do  not  release  key-press  until 

"--CAL--  "  is  displayed).  Then  the  required  calibration  weight  is 

displayed  (ex.  3000.0g  means  3kg  weight).Put  the  corresponding 

calibration  weights  on  the  pan,  the  scale  will  display  "-  -  -  -  -  -"  for  a 

while , and then display the value of weight. Then remove the calibration 

weight, the scale  will  display  "-  -  -  -  -  -  "  and then  display  "0.0"  to  end 

calibration. 

D. ERROR INDICATION 

Error information 

Explanation 

Solution 

Underline----------- 

The weight is negative    Recalibrate 

Hyphen----------- 

Overload 

Recalibrate 

Not accurate weighing 

and cannot calibrate 

Load cell damaged 

Change load cell 

Line------ and display      Battery finishes 

Recharge battery 

 

 

 

Reviews: