Externally taked drugs L-dopa, dopamine,
methyl-dopa, acetaminophen, and
ibuprofen will not interfere with SD
CodeFree™ blood glucose results when
taken at therapeutic concentrations.
•
The SD CodeFree™ System is not designed
to be a substitute for pathology laboratory
equipment and should not be used for the
diagnosis of diabetes.
•
Use only fresh capillary blood. Do not use
serum or plasma or venous whole blood.
•
Do not use the SD CodeFree™ meter to
test neonates. It has not been validated for
neonatal use.
•
Always insert the test strip into the meter
first, and then prick the finger.
•
Interferences : The following compounds,
elevated levels of ascorbic acid, uric
acid, acetaminophen, total bilirubin and
triglycerides may affect results.
Compound
Limitation
Ascorbic acid
> 4 mg/dL
Uric acid
> 9 mg/dL
Acetaminophen
> 6 mg/dL
Total bilirubin
> 40 mg/dL
Triglycerides
> 1026 mg/dL