SD GlucoNavii® Voice N BLOOD GLUCOSE MONITORING SYSTEM
C
on
tr
ol solution displa
y
33
3)
The control solution result appears on the screen in just
5 seconds.
4)
Compare control solution result with the range printed
on the test strip container.
If the results are not within the control range printed on
the test strip container, then the meter and strips may
not be working properly. Repeat the control solution
test.
5)
Remove the used test strip for control solution from
the meter and discard it.
mg/dL
mmol/L
VOICE N
NOTE
The control solution range printed on the test strip
container is for SD GlucoNavii® Control Solution only.
It is not a recommended range for your blood
glucose level.
[This is an example. Refer to the ranges on your test
strip container.]
Control Range
Level M
Level H
90-140 mg/dL
170-240 mg/dL
5.0-7.8 mmol/L
9.4-13.3 mmol/L
mg/dL
mmol/L
[Example]