Using the LEDs to Set the Resistance Alarm Level

Resistance 

(Ohms)

LED

1

2

3

4

5

6

7

8

9

10

12

14

16

18

20

Calibration

Equipment Required

For this test, the Wrist Strap and Ground Monitor should 

be disconnected from the monitored ground.
Use the SCS CTE701 Workstation Monitor Checker.  

Connect the ground input of the checker to the ground 

terminal of the Wrist Strap and Ground Monitor.  Set 

the checker to a proper resistance alarm level (see user 

guide 

TB-9031

).  Connect the red tip of the checker 

to the GND1 terminal.  Press the Fail button on the 

checker.  The red LED of the corresponding ground of 

the monitor should remain on.  Press the Pass button 

on the checker – the corresponding light on the monitor 

should be green.
The checker verifies the proper operation of dual wrist 

strap monitoring.  Connect a 3.5 mm test cable to both 

the checker and to the operator jack of the monitor.  At 

this point the monitor should indicate failure.  Depress 

the Pass button on the wrist strap section.  The monitor 

should indicate a good connection.  While pressing the 

Pass button, press body voltage “high.”  At this point the 

Wrist Strap LED would be blinking red.