40
CHAPTER 5: MAINTENANCE
087-0038 REV L
SCOTT SAFETY
4
If display indicates IS AIR CLEAR?, and atmosphere has been verified free, press
the Right Operation Button to start the Zero Calibration.
5
Wait for device to complete Zero Calibration. If passed, the LCD displays APPLY
GAS. See
.
Figure 26
Zero Calibration Display - Passed
6
Wait for device to complete Zero calibration. If failed, the LCD displays FAILED
and the appropriate failing sensor(s). See
.
Figure 27
Zero Calibration Display - Failed
7
If Span calibration is desired, continue to See
“Span Calibration” on page 40.
8
If Span Calibration is not desired, either press the Left Operation Button to exit, or
allow 80 second timer to expire.
9
When the LCD displays CAL STOPPED, press either Operation Button (Left or
Right) to exit calibration mode.
Span Calibration
This section describes how to perform Span Calibration.
Warning:
Read, understand and follow the instructions, including labels and MSDS
provided by the manufacturer of the calibration gas. Failure to do so may result in
injury or death.
For precise K-Factors when performing a Span Calibration of a combustible sensor.
See
Summary of Contents for Protege
Page 1: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Page 2: ......
Page 3: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Page 10: ...X LIST OF TABLES 087 0038 REV L SCOTT SAFETY ...
Page 44: ...20 CHAPTER 2 INSTALLATION 087 0038 REV L SCOTT SAFETY ...
Page 60: ...36 CHAPTER 4 OPERATION 087 0038 REV L SCOTT SAFETY ...
Page 76: ...52 CHAPTER 5 MAINTENANCE 087 0038 REV L SCOTT SAFETY ...
Page 80: ...56 APPENDIX A SPECIFICATIONS 087 0038 REV L SCOTT SAFETY ...
Page 86: ...62 APPENDIX B QUICK REFERENCE 087 0038 REV L SCOTT SAFETY ...
Page 90: ...66 APPENDIX C SENSOR INFORMATION 087 0038 REV L SCOTT SAFETY ...
Page 102: ...78 INDEX 087 0038 REV L SCOTT SAFETY ...
Page 103: ......