CLEANING INSTRUCTIONS
The following instructions are specific to our
REUSABLE
laryngoscope blades and handles. These
instructions are intended for use only by authorized personnel with the required knowledge and training.
PRODUCTS MARKED “ / DISPOSABLE / SINGLE PATIENT USE / DO NOT AUTOCLAVE”
SHOULD NOT BE REPROCESSED.
WARNINGS
1. Follow instructions and warnings as issued by the manufacturer of any decontaminants, disinfectants
and cleaning agents used. Wherever possible avoid use of mineral acids and harsh, abrasive agents.
2. Do not use chemical sterilants with caustic ingredients such as - surgical scrub solution, povidone-
iodine solution, bleach, peroxide solution, virox 3 and sporox. Do not use garment or surface
disinfectants.
3. Do not autoclave with sub standard stainless steel surgical instrument as this may cause a reaction
and lead to rust or discoloration.
4. Ultrasonic and plasma cleaning are not permitted, as this will cause damage to the fiber optic light
guide and lamp.
5. Do not flash autoclave or dry heat blade or handle, this will invalidate the warranty .
6. Our laryngoscopes are suitable for sterilization by Ethylene Oxide (ETO), Sterris and Sterrad
sterilization procedures.
When reprocessing medical devices, handle with care, wearing protective clothing and face
visors or goggles.
BATTERIES AND LAMPS MUST BE REMOVED FROM THE HANDLE / BLADE
PRIOR TO CLEANING.
LIMITATIONS ON REPROCESSING:
Repeated autoclaving, especially for extended periods will affect
the fiber optic light guide ends which may lead to a reduction in light output.
FROM POINT OF USE:
If possible, soiled blades should be placed in a holding solution (combined
disinfectant / enzyme solution) immediately after use and prior to cleaning.
PREPARATION FOR DECONTAMINATION:
Reprocess all blades as soon as it is reasonably practical
following use. We recommend disassembling the fiber optic light guide from Modular blades for separate
cleaning.
CLEANING-AUTOMATED:
Use only validated washer-disinfector machines and cleaning agents,
following the manufacturer's instruction for use, warnings and recommended cycles. Load blade and
handle carefully, ensure the fiber optic light guides are not liable to damage during loading procedure.
Bulbs should not be cleaned in these machines.
CLEANING-MANUAL:
Use a sink dedicated for instrument cleaning (not used for hand washing), rinse
excess soil from blades & handles (water temp <35°C) Keeping the blades submerged, with a brush,
apply cleaning solution to all surfaces. Pay particular attention to underside of the blade and heel. Always
brush away from the body. Do not use a wire brush. Take great care not to brush against the Fibre ends in
the blades to avoid scratching or damage. Rinse blade thoroughly with clean water, so that the water
reaches all parts of blade, then carefully hand dry or use a drying cabinet.
DISINFECTION:
To disinfect Cidex® OPA or 2.4% Glutaraldehyde solution may be used. Always follow
the manufacturers instructions. After aplication, dry with a clean cloth or filtered air.
LAMP CLEANING:
Lamps should only be cleaned manually. Soak surgical grade cotton swab in
decontaminant and apply to wipe off lamp gently. Repeat this procedure twice on every cleaning cycle.
Carefully hand dry lamp before placing back in position.
Replace batteries with the size and type specified in your product's User Manual. Remove all batteries
from the product at the same time.
Charge rechargeable batteries with the instructions and correct charging equipment supplied with
your product.
Store batteries in a cool, dry place at normal room temperature, Remove batteries from
handles that will be stored unused for extended period of time.
Dispose off batteries according to the local laws and regulations of your region. Some batteries may
be recycled, and may be accepted for disposal at your local recycling centre. If you are not able to
identify the applicable rules in your area, please check the instructions of the battery manufacturer.
Contact supplier
if you have questions about proper handling of batteries. Batteries may present a risk
of fire, explosion, or chemical burn if mistreated.
Do not disassemble, puncture, modify, drop, throw, or cause other unnecessary shocks to batteries.
Do not dispose off batteries in a fire or trash incinerator, or leave batteries in hot places such as an
automobile under direct sunlight.
Do not store batteries near an oven, stove, or other heat source.
Do not connect batteries directly to an electrical source, such as a building outlet or automobile power-
point.
Do not place batteries into a microwave oven, or into any other high-pressure container.
Do not immerse batteries in water or otherwise get them wet.
Do not short circuit batteries; do not carry loose batteries in a pocket or purse with other metal objects,
which may inadvertently cause a battery to short circuit.
Do not use batteries, or charge rechargeable batteries, that appear to be leaking, discoloured & rusty,
deformed; emit an odour; or are otherwise abnormal.
Do not touch leaking batteries directly; wear protective material to remove the batteries and dispose
off them properly immediately.
Do not mix old/used with new batteries, or install used batteries in other equipment.
Do not charge non-rechargeable batteries, or use unspecified charging instructions or equipment for
rechargeable batteries.
Do not continue to charge batteries beyond the specified time.
Do not give batteries to young children, or store batteries where children may access them.
Do not put batteries in mouth. If swallowed please contact physician or local poison control centre
immediately.
Warning
Power outputs from some rechargeable batteries can fall rapidly during use, resulting in rapid
failure of illumination.
Recycling is the best option from both an economic and environmental standpoint for most
batteries. Check with companies to make sure that they will take back old batteries when new
ones are bought. If batteries cannot be recycled (with the exception of alkaline batteries), make
sure they are tagged for collection by designated personal.
IFU/multiple/2.0/rev1.0/04/2017
1
4
5
2
3
‘AA
’
‘AA
’
‘AA
’
‘AA
’
1
3
2
‘C’
‘C’
BATTERY SIZE & INSTALLATION
1
3
2
‘AA
’
‘AA
’
STERILIZATION:
To sterilize, use a validated vacuum
autoclave operating at 134°-137°C @ 2.25 Bar for a
minimum holding time of 3 minutes. Always follow the
instructions of machine manufacturer. When sterilizing
multiple instruments in one autoclave cycle, ensure that
sterilizer manufacturers stated maximum load is not
exceeded. Ensure blades are dry before sterilization, if they
cannot be dried prior to sterilization, then use distilled / de-
ionized water in the final-rinse stage of cleaning.
REASSEMBLY:
If the blade or handle has been
disassembled (as may be the case with Modular fiber optic
blades), reassemble and ensure the device works properly.
Check the fiber optic light guide fits correctly. Do not apply
excessive force. Lamps should be placed back in their original housing securely.
INSPECTION AND FUNCTION TESTING:
Visually inspect and check all lamps and fibre optic light
guides for damage and wear. To check functionality, connect the laryngoscope blade to the handle and
pull upward to the ON position. If the unit fails to light or flickers, check the lamp and component fittings
battery.
PACKAGING:
All blades and handle should be packed following local protocols.
STORAGE:
Ensure blades and handles are dry before storage and stored one in dry, clean condition at
an ambient room temperature.
ADDITIONAL INFORMATION:
Other forms of cleaning (alkaline and neutral) and sterilization (Cidex
OPA, Ethylene oxide Up to 65° C for anesthetic devices) are permitted. However, always follow the
instructions for use as issued by the manufacturer and always consult with them if in any doubt over the
suitability of any process used.
It is the responsibility of the reprocessor to ensure that the reprocessing has actually been
performed using appropriate equipment, materials and personnel and the desired results have
been achieved. This requires validation and routine monitoring of the process. Likewise any
deviation by reprocessor from the instruction provided, must be properly evaluated for
effectiveness and potential adverse consequences.
The above listed sterilization guidelines, provided by manufacturer, are intended as suggestions
only. They are based on procedures compatible with specific materials. Sterilization must be
performed to approved hospital protocol. Manufacturer cannot guarantee that any of the
recommended methods will guarantee sterility. This must be validated by the hospital and / or
sterilization equipment manufacturer.
BATTERY INFORMATION :
Remove the screw cap to insert appropriate batteries into handle shaft. For medium handles use two 'C'
size batteries, for small and stubby handles use two 'AA' size batteries. When inserting batteries the ‘+ve’
terminal must be inserted into the tube first on both of the batteries so that the ‘–ve’ terminal is facing the
open end. For stubby handles - batteries should be inserted following the negative ‘–ve’ and ‘+ve’
terminals as shown on plastic battery housing provided in all stubby handles. Replace the base cap and
ensure it is securely tightened.
Alkaline Batteries:
are commonly referred to as disposable or non– rechargeable batteries.
Rechargeable Batteries:
typically include Nickel – Cadmium (Ni-Cad) or Lithium Ion Batteries.
DO NOT AUTOCLAVE BATTERIES UNDER ANY CIRCUMSTANCES.
WE RECOMMEND USING ONLY “DURACELL” BATTERIES IN OUR LARYNGOSCOPE HANDLES.
LAMP REMOVAL IN LED HANDLES