Scinomix TS-1023 Instructions Manual Download Page 6

TS-1023 Ribbon Tension

Adjustment

(VX2/VXL)

Document Owner:

Service Department

Effective Date:

AUG-22-2022

Page 6 of 7

a. If experiencing ribbon wrinkling, moving the pin up may help reduce

wrinkling.

b. Use a 2.5mm Allen wrench to loosen the set screw in the face of the top

guide pin, then slide the pin up.

9.

Check angle of printhead guide:

While this does not happen often and typically

the checks before this fix it, it may be necessary to change the angle on the

printhead guide rail. There is a center screw on the back side of the printhead

that when loosened, adjusts the pitch of the printhead guide. At the top and

bottom of this guide, there is a socket head screw. The one at the top is used as

a stop (notice the larger hole that it is positioned in. Ideally this screw is centered

in this hole, but may need to be brought slightly towards the front.  Picture below.

10.

Test by printing multiple labels

. Ribbon wrinkling should be minimal on the

ribbon/label back feed after each printed label.

Figure 6.

11.

Check the Ribbon Source Rewind

. After the ribbon is properly installed, close

the print head and confirm the ribbon is taut.

a. Turn the source reel clockwise ½ turn and release.

b. Confirm that the internal spring rewinds the source reel and tensions the

ribbon. (If not, see

TS-1049 Installing the Shim Washer into a SQUIX

Printer

).

File Location: Google Drive

Revision: B

Approved Date: AUG-22-2022 Approved by: Service

Summary of Contents for TS-1023

Page 1: ...white lines through the print on the label ribbon wrinkles Required Material and Equipment 2 5mm Allen wrench for Squix printer 2mm Allen wrench for A2 printer Intended Audience Scinomix End User Int...

Page 2: ...printer from front of instrument 2 Ribbon Attached to Core at Bottom Confirm the ribbon is wrapped around the ribbon core counter clockwise Make sure a The bottom edge of the ribbon is touching the bo...

Page 3: ...n the base of the ribbon tension bar a The graduations and the line on the base of the ribbon tension bar are used to determine relative position b Typically the bar is set between the center line and...

Page 4: ...tive position b Make sure that the scribed line on the wrinkle bar is slightly to the positive side of the centerline Picture is below 6 Check Adjust Top and Bottom Guide Pin Tension The core of the g...

Page 5: ...same position See Figure 6 below Figure 5 Figure 6 7 Check the ribbon roll is not over splayed If the ribbon roll is over tightened to the point the reel on which it sits is at an angle this increases...

Page 6: ...e top and bottom of this guide there is a socket head screw The one at the top is used as a stop notice the larger hole that it is positioned in Ideally this screw is centered in this hole but may nee...

Page 7: ...al assistance please call our service department at 314 298 9800 or email service scinomix com Definitions N A Revision History Version Change Effective Date Approved by A NEW APR 24 2020 B Petti B Co...