Page 1 of 7

Purpose

This document provides instructions for adjusting the ribbon tension on the Sci-Print

VX2/VXL.

Intended Use

If experiencing issues with the ribbon ripping, or diagonal white lines through the print on

the label (ribbon wrinkles).

Required Material and Equipment

● 2.5mm Allen wrench (for Squix printer)

● 2mm Allen wrench (for A2 printer)

Intended Audience

Scinomix End User

Intended Audience

Field Service Representative from Scinomix or approved vendors.

Prerequisites

N/A

Procedure

1.

Ribbon Core.

Make sure the ribbon core is tightened against the bottom of the

ribbon take-up reel.

File Location: Google Drive

Revision: B

Approved Date: AUG-22-2022 Approved by: Service

Summary of Contents for TS-1023

Page 1: ...white lines through the print on the label ribbon wrinkles Required Material and Equipment 2 5mm Allen wrench for Squix printer 2mm Allen wrench for A2 printer Intended Audience Scinomix End User Int...

Page 2: ...printer from front of instrument 2 Ribbon Attached to Core at Bottom Confirm the ribbon is wrapped around the ribbon core counter clockwise Make sure a The bottom edge of the ribbon is touching the bo...

Page 3: ...n the base of the ribbon tension bar a The graduations and the line on the base of the ribbon tension bar are used to determine relative position b Typically the bar is set between the center line and...

Page 4: ...tive position b Make sure that the scribed line on the wrinkle bar is slightly to the positive side of the centerline Picture is below 6 Check Adjust Top and Bottom Guide Pin Tension The core of the g...

Page 5: ...same position See Figure 6 below Figure 5 Figure 6 7 Check the ribbon roll is not over splayed If the ribbon roll is over tightened to the point the reel on which it sits is at an angle this increases...

Page 6: ...e top and bottom of this guide there is a socket head screw The one at the top is used as a stop notice the larger hole that it is positioned in Ideally this screw is centered in this hole but may nee...

Page 7: ...al assistance please call our service department at 314 298 9800 or email service scinomix com Definitions N A Revision History Version Change Effective Date Approved by A NEW APR 24 2020 B Petti B Co...

Reviews: