Rotary Evaporator 

 

safety of the heating bath. 

 

The  heating  bath’s  safety  temperature  shall  at  least 

25  °C  lower  than  the  ignition  point  of  the  medium 

being handled. 

 

Please  don’t  operate  or  use  this  instrument  in  an 
explosive  environment  or  underwater  or  use  this 

instrument to handle any dangerous substance. 

 

This  instrument  is  only  suitable  for  the  medium  that 

does  not  react  with  the  energy  generated  in  the 

handling  process  in  a  way  that  might  cause  danger. 

Meanwhile,  the  substance  being  handled  may  not 

produce  danger  due  to  energy  generated  in  any  other 

manner such as illumination reaction. 

 

The  instrument  must  be  monitored  all  the  time  when 

operating the instrument. 

 

Please  don’t  operate  this  instrument  under 
overpressure.   

 

To  ensure  the  effective  condensation  of  the  motor 

system,  please  do  not  block  the  vent  of  the  motor 

system. 

 

Only  use  of  the  accessories  in  the  list  of  “optional 
accessories” can guarantee the operational safety. 

 

Please  refer  to  the  operating  instructions  for  heating 

kettle. 

 

Please  refer  to  the  operating  instructions  for  optional 

accessories. 

 

Please  use  this  instrument  in  a  closed  fume  hood  or 

other suitable protective device. 

 

Select the distilling device according to the amount to 

be  distilled  and  the  distillation  type.  The  condenser 

must  work  normally.  Please  monitor  the  flow  rate  of 

the coolant at the outlet of the cooler. 

 

To  prevent  pressure  buildup,  the  glass  assembly  shall 

be kept vented when operating under normal pressure, 

such as keeping the condenser open. 

 

Please  note  that  gases,  vapor  or  other  substances  can 

pose danger through overflow from the upper opening 

of  the  condensation  tube.  To  minimize  or  avoid  such 

danger,  please  take  suitable  measures,  such  as 

connecting  the  cooling  tube  and  scrubbing  bottle  or 

other effective extraction devices downwards. 

 

The glass evaporating flask may not be heated on one 

side. The  rotating evaporating  flask  must rotate at the 

heating phase. 

 

The  glass  assembly’  design  pressure  tolerance  is 
10mbar. The venting valve (see the part of “Testing”) 

Summary of Contents for SCI100-S

Page 1: ...a at t to o or r r Operation Instructions LED Digital Rotary Evaporator Please carefully read the Instructions and use the product safely under the direction of the Instructions If any information of...

Page 2: ...he use of the Manual the company will not be liable for this at all The manual is a pure technical document free of any implication or oblique hint of any third party Moreover we won t be liable for a...

Page 3: ...cation 5 3 Inspection 6 3 1 Open package inspection 6 3 2 Package list 6 4 Control and display 6 4 1 Control 6 4 2 Display 9 5 Installation and testing 11 5 1 Installation 11 5 2 Testing 14 6 Fault di...

Page 4: ...nty card Description of problem phenomenon The method and operating steps you have undertaken to resolve the problem Your contact details including telephone number fax number and email address Warran...

Page 5: ...obiological medium might cause damage to the operator Place the instrument on a stable clean antiskid dry and fireproof table surface Make sure there is sufficient space above the instrument and the g...

Page 6: ...rantee the operational safety Please refer to the operating instructions for heating kettle Please refer to the operating instructions for optional accessories Please use this instrument in a closed f...

Page 7: ...flask and glass assembly might become hot in the operation process and remain hot for a while after stop of the operation Before further operation please let each part cool down Note be careful to av...

Page 8: ...es are protected from extrusion and collision Only professional repairman may open the instrument 2 Scope of application This instrument is designed for such application environments as schools labs a...

Page 9: ...e contact supplier or manufacturer Caution 1 If you find any visible damage on the product please don t connect the product to power supply 3 2 Package list Name Qty Main unit 1 Heating bath 1 Power c...

Page 10: ...Rotary Evaporator 7 Fig 2 Fig 3 Main unit three in one switch Heating bath power interface Heating bath power switch...

Page 11: ...erse rotation interval time Turning the adjustment knob can set the intermittent left and right rotation time within the range of 1 999 sec When set to 0 sec the instrument rotates forward only Adjust...

Page 12: ...es the real time speed Fig 6 This interface indicates the real time speed Turn the adjustment knob to set the speed and press the adjustment knob to start the instrument The rotation speed can be chan...

Page 13: ...get temperature of the heating bath After setting by pressing the knob the heating indicator is on and the heating process begins In the heating process the target temperature can be changed by turnin...

Page 14: ...Fig 11 Place the main unit on a flat tabletop and squeeze the handle switch a to lift it to the highest position Fig 12 Tighten the screw b by hand adjust the angle of the motor head and tighten the...

Page 15: ...mount the evaporating tube in place and rotate the locking device d clockwise for locking Mount the seal ring Fig 13 and Fig 14 Fig 14 Fig 15 Mount the condenser f Fig 15 and Fig 16 Slide the locking...

Page 16: ...ry Evaporator 13 Fig 16 Fig 17 Mount the fixing device for the vertical condensation glass assembly Fig 17 Fig 18 Piping connection Fig 18 f Condensate inlet Vacuum tube interface g h g Condensate out...

Page 17: ...90 under the water bath mode Press the temperature adjustment knob to start heating Squeeze the handle switch to place the evaporating flask into the heating bath Press the adjustment knob to start ro...

Page 18: ...supply until all surfaces are dry If liquid or moist solid enters the product please disconnect the power supply quickly and leave off and contact the manufacturer supplier for more advice Keep the p...

Page 19: ...ower W 45 1010 Motor DC brushless Speed range rpm 20 200 Display LED Temperature range C Room temperature 180 Temperature control accuracy C 1 oil 3 Lifting method Manual Lifting travel mm 110 Forward...

Page 20: ...Rotary Evaporator 17 External dimension W x D x H mm 440 320 450 300 300 240 Weight kg 7 3 Allowable ambient temperature C 5 40 Allowable RH 80 DIN EN60529 protection class IP20 USB Yes...

Page 21: ...SCILOGEX LLC 1275 Cromwell Ave Suite C6 Rocky Hill CT 06067 USA Tel 1 860 436 9221 Fax 1 860 436 9745 info scilogex com www scilogex com...

Reviews: