![Scilogex DM 0412 User Manual Download Page 17](http://html1.mh-extra.com/html/scilogex/dm-0412/dm-0412_user-manual_1217644017.webp)
14
11.1.2
Available rotors and adapters
Rotor
type
Tubes
Tubes
Per
Rotor
Dimension
(
Ф
×L
mm)
Adapters
Maximum
speed
(
rpm
)
Radius
cm
Maximum
RCF
(×g
)
A12-10P
15ml con
6
17×120
4500
11
2490
1.5-5ml vacu
12
13×82
A10P15
mat
4500 9.8
2218
4-7 ml vacu
12
13×106
A10P15
4500
11 2490
16×75
A10P15
Mat
9.8 2218
8.5-10 ml vacu
12
16×107
A10P15
4500
11
2490
2.7-3
(EU) ml
collection tube
12 11×66
A10P15
mat
4500 9.8
2218
7.5-8.2
(EU) ml
collection tube
12 15×92
A10P15
4500
11 2490
Table 11.1 Rotors and adapters
11.1.3 Notice
z
The centrifuge rotor can separate samples with a density lower than 2.0g/ml. If the samples density is
over 2.0g/ml, please calculate allowable speed depending on the following formula.
Allow Speed (rpm)= Maximum speed×(2.0(g/ml)/Sample density (g/ml))
1/2
11.1.4 Autoclaving
A12-10P rotor is made of plastic, cannot be high-pressure sterilization and UV irradiation, only ordinary
sterilization can be used.