Calibration Procedure
Step 1)
Make sure you have a certified reference thermometer that is at least
twice the accuracy of the thermometer to be recalibrated. Eg. RD0370R
https://www.scigiene.com/Scigiene-Reference-Thermometer-NIST-Traceable
Step 2)
Insert temperature probe into
https://www.scigiene.com/SCCAL-2-
https://www.scigiene.com/SCCAL-9-Calibrator
https://www.scigiene.com/TempCalibrator140
at 0°C in order to calibrate
precisely, and follow the steps in the calibration setup instructions. NOTE regular
ice bath methods are not accurate enough to recalibrate these devices in order to
meet stated specifications.
Temperature Meter Calibration Setup
1.
Below table diagram shows each button function when user enter into
calibration mode.
Note:
Turn the Power “OFF” before attempting following SETUP.
Setup mode is cancelled during below procedure if “POWER” button is pressed.