scigiene RD0370 Instruction Manual Download Page 1

 

Digital RTD Thermometer 

Part# RD0370

 

 

Instruction Manual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for RD0370

Page 1: ...Digital RTD Thermometer Part RD0370 Instruction Manual...

Page 2: ...cifications 3 4 Instrument Description 5 6 Operating Instructions 7 9 Temperature Measurement 10 Battery Replacement 11 Calibration Procedure 12 Temperature Meter Calibration Setup 12 14 Product Maint...

Page 3: ...ith LED backlight C F switchable Min Max Avg functions Relative mode and Data hold Low battery indication with auto power off Includes piercing type temperature probe Pt 100 IEC751 Class A accuracy Id...

Page 4: ...the following tolerance into the accuracy specification 0 01 of reading 0 03 C 0 01 of reading 0 06 F The basic accuracy Specification does not include the error of the probe refer to the probe accura...

Page 5: ...Instrument Description 1 Pt type temperature sensor connector 2 LCD display 3 Power ON OFF button 4 HOLD button 5 MAX MIN Average control button 6 Back light button 7 Relative readout button 5...

Page 6: ...tion Pt xxx Platinum Type Indication HOLD This indicates that the display data is being held MAX The Maximum value is being displayed MIN The Minimum value is being displayed AVG The Average value is...

Page 7: ...ssing C F button and vice versa back to Celsius Data Hold Operation To hold the present reading and keep it on the display press the HOLD button When the hold data is no longer needed release the data...

Page 8: ...he last 8 readings from the memory it will update with every new reading When the MAX symbol is displayed the Maximum value is shown on the display Press AVG MAX MIN again and the NIN symbol is on the...

Page 9: ...disabled Low Battery This meter indicates the battery power level directly on the display Use this reference guide to establish battery power level Battery power level is FULL you can still take meas...

Page 10: ...t the end of the metal sheath of the sheath type temperature probe To accurately test internal temperatures insert the probe to a distance of at least 15 times the diameter of the sheath directly into...

Page 11: ...into the instrument 3 Replace cover properly onto the meter and tighten screws The unit s back cover is fitted with rubber rings After replacing the batteries check that the rubber rings are properly...

Page 12: ...SCCAL 9 Calibrator or https www scigiene com TempCalibrator140 at 0 C in order to calibrate precisely and follow the steps in the calibration setup instructions NOTE regular ice bath methods are not a...

Page 13: ...en PT probe type Pt100 Pt500 Pt1000 is circulate Press C F button to confirm selection Press and hold Power Light buttons for 2 seconds to enter setup mode Press REL button to enter calibration mode E...

Page 14: ...2 seconds to enter setup mode Press REL button to enter calibration mode Enter PT selection first Press C F button to confirm PT selection Press Light button to enter recall mode Press C F button to...

Page 15: ...at contain carbon or benzenes alcohol or anything similar to clean the meter as these substances will damage the surface of the meter Do not use tools with sharp edges screwdrivers metal brushes or an...