Table D-1 NanoSpray III Head—Positive Mode
Range
Typical Value
Parameter
50 nL/min to 2000 nL/min
500 nL/min
Flow rate
2 mm to 5 mm
5 mm
Sprayer distance from curtain plate
1000 V to 3000 V
2300 V
IonSpray
TM
Voltage (IS) or IonSpray
TM
Voltage Floating (ISVF)
1 to 20
6
Nebulizer gas (GS1)
15 to 30
20 (Triple quadrupole and
QTRAP
®
systems)
Curtain Gas
TM
flow (CUR)
25 (TripleTOF
®
systems)
50 °C to 150 °C
75 °C
Interface heater temperature (IHT)
Table D-2 NanoSpray III Head—Negative Mode
Range
Typical Value
Parameter
50 nL/min to 2000 nL/min
500 nL/min
Flow rate
2 mm to 5 mm
5 mm
Sprayer distance from curtain plate
–1000 V to –2800 V
-1900 V
Ionspray Voltage (IS) or Ionspray
Voltage Floating (ISVF)
1 to 50
20
Nebulizer gas (GS1)
15 to 30
20 (Triple quadrupole and
QTRAP
®
systems)
Curtain Gas flow (CUR)
25 (TripleTOF
®
systems)
50 °C to 150 °C
75 °C
Interface heater temperature (IHT)