Mass Spectrometer Troubleshooting Tips
Corrective Action
Possible Cause
Symptom
a. Loosen the plug in the port so that the
inlet is not sealed. (Do not remove the
plug.) Refer to
b. Set GS2 to 0 in the acquisition method.
The Gas 2 inlet port on the ion source is
blocked and Ion Source Gas 2 (GS2) is set to
a value other than 0 in the acquisition
method.
Gas 2 Over Pressure
error occurs and the
electronics turn off
Figure 7-1 Gas 2 Port