Item
Reservoir bottle
A
Tags on suction tubing
a
Degasser ports
B
2. Put the suction filter and the tube of the mobile phase port to be used in the reservoir bottle.
Tags on the suction tubing identify the degasser ports to which the tubing should be connected. Consult the
tags and install each tube in the reservoir bottle.
Table 4-1 Tubing Tags
Port Name
Tag Label
Rinse solution for autosampler
Rinse
Mobile phase A
SOLVENT: A
Mobile phase B
SOLVENT: B
Mobile phase C
SOLVENT: C
Mobile phase D
SOLVENT: D
3. Install the reservoir bottle on the reservoir tray.
Perform an Autopurge
The mobile phase ports and the rinsing port of the autosampler are automatically purged.
• Fill the whole flow line with mobile phase before installing the column.
• Purge mobile phase ports that are not used with a liquid that is equivalent to the mobile phase or that does
not affect analysis results. Be careful not to allow air bubbles to remain in the tubes of the mobile phase ports
that are not used.
• When the
Rinse solution aspiration rate
in the autosampler settings is 35 µL/sec, the average flow rate
of rinse solution pumped by the measuring pump is approximately 1.0 mL/min. To completely replace liquid
in the flow line with new rinse solution, perform an autopurge for 25 minutes.