Scientemp 85-6.8A Operating And Installation Manual Download Page 3

freezer improperly may jeopardize safety or cause undue 
stress or damage to the refrigeration compressors. 

Avoid opening the door for extended periods of time since 
chamber air will escape rapidly.  Room air, which is higher 
in humidity, will replace chamber air and may cause frost to 
develop in the chamber more rapidly. 

4. CABINET INSTALLATION START-UP 

Once the cabinet has been placed in its permanent location 
and the proper power and grounding have been provided, the 
following items must be checked or completed: 

1.

 

Check for traces of oil in the compressor area, 
which could mean a broken or leaking refrigeration 
line. 
 

*CAUTION:

 

UNDER NO CIRCUMSTANCES 

SHOULD THE COMPRESSOR BE STARTED 
WHEN OIL IS PRESENT UNTIL INSPECTED 
BY A SERVICE TECHNICIAN. 

2.

 

Inspect the factory wiring for terminals that might 
have vibrated loose in shipping.  Tighten all screw-
type terminals. 
 

3.

 

Check the refrigeration lines to insure that they do 
not rub and no damage was done during shipping of 
the cabinet. 

 

4.

 

Check fan blade(s) for “free” operation. 

 

5.

 

When freezer is plugged in and the compressor 
starts, the voltage should be checked at the 
compressor terminals to determine if there is proper 
voltage to the compressor.  The voltage should not 
exceed the 10% above or below the rated 
compressor voltage. 

 

After reading and completing the Safety Considerations, 
Installation and Operation section of this manual, you are 
now ready to plug in the freezer.  

1.

 

Check for obstructions with fan blades.  
 

2.

 

Cabinet should not be loaded with product until the 
cabinet has operated for 24 hours at the correct 
temperature. 

 

Do not exceed the electrical and temperature ratings printed 
on the data plate located on the upper left backside of the 
cabinet. 

 

*CAUTION:

  Improper operation of the cabinet 

could result in dangerous conditions.  Follow all instructions 
and operate within the design limits noted on the data plate. 

Now that your cabinet has been trouble checked, run the 
cabinet down to storage temperature before loading product. 

 

4.1 Lid Seal 

If the lid gasket does not seal along the front edge, relieve 
the compression of the gasket along the back edge by 
loosening the hinge screws where they attach to the cabinet.  
Raise the lid slightly and then retighten the screws. 

 

4.2  Interior Compartment Temperature 

The temperature is controlled by a CAL 9500 digital 
temperature controller. When you first power up the freezer,  
the high stage compressor and fan will start and operate for 3 
to 6 minutes before the low stage compressor starts.  Pull 
down to minimum temperature will take from 3 to 6 hours, 
depending on ambient temperature and size of cabinet.  The 
high stage compressor runs 100% of the time.  The low stage 
compressor will cycle on and off to maintain the temperature 
of the controller setting.  Set the control to the desired set 
point. To view the set point temperature, press the * and the 
UP arrow keys together. To decrease the set point 
temperature, press the * and the DOWN arrows keys 
together. 

Freezers are shipped from the factory with the indicating 
digital control programmed for proper operation. All but the 
temperature settings may be adjusted. The temperature 
sensor is a type 

“T”

 thermocouple located on the front wall 

of the inner-liner. 

4.3

 

Sonalarm 

The Sonalarm should be kept in the “OFF” position until the 
cabinet temperature has reached the appropriate setting.  
When the temperature has been reached, the temperature 
alarm should be switched to the “ON” position and is now 
ready to operate.  It will monitor “temperature rise” and 
“power-failure”.  The control is factory set to operate the 
Sonalarm at approximately 10

°

F (6

°

C) above the set 

temperature.  No further adjustment is necessary unless a 
spread other than this is desired.  In the event a temperature 
spread of more or less than factory set is desired, refer to 
your digital temperature control manual.  
 

5 ELECTRICAL CONNECTIONS 

5.1 Wiring 

 

The cabinet must be plugged into a properly grounded 
electrical outlet.  Electrical connection should be made in 
compliance with local codes.  

We suggest a separate circuit with a fused disconnect be 
installed. To avoid the possibility of an accidental disconnect 

Summary of Contents for 85-6.8A

Page 1: ...517 263 5492 Website www scientemp com OPERATING AND INSTALLATION MANUAL MODEL 85 6 8A This cabinet has passed the QUALITY CONTROL INSPECTION and meets the high standards at Scientemp Corp This inspection includes the complete refrigeration system cabinet construction and finish ...

Page 2: ...sing or maintaining this product please be sure to read this manual carefully Failure to follow these instructions may cause this product to malfunction which could result in injury or damage Below are important safety precautions that apply to this product Use this product only in the manner described in the product literature and in this manual Before using it verify that this product is suitabl...

Page 3: ...rature before loading product 4 1 Lid Seal If the lid gasket does not seal along the front edge relieve the compression of the gasket along the back edge by loosening the hinge screws where they attach to the cabinet Raise the lid slightly and then retighten the screws 4 2 Interior Compartment Temperature The temperature is controlled by a CAL 9500 digital temperature controller When you first pow...

Page 4: ...you use a freezer or handle a frozen product that is at temperatures below 0 Centigrade or 32 Fahrenheit 6 1 When Skin Freezes When your skin is exposed to subfreezing temperatures for an extended amount of time it can freeze Your blood vessels constrict in response to dropping temperatures This reduces the flow of blood and therefore the amount of oxygen to the tissues When water in these tissues...

Page 5: ...d by a small amber light which is lit whenever power is available 2 Temperature rise is indicated by alternate flashing lights accompanied by a repeating audible signal in the form of beep beep beep 3 Power failure is indicated by alternate flashing lights accompanied by a repeating audible signal with the power on light off The Sonalarm is powered by four 4 D sizes 1 volt dry cell batteries It is...

Page 6: ...ve white plastic breaker by pulling up and out on edge next to inner liner 4 Lift insulation and remove hard board sub breaker 5 Remove instrument panel 4 screws and lay in front of freezer face down 6 Remove T C wires from terminals 7 Remove rubber grommets from cabinet bottom inner liner 8 Remove clips over T C 9 Tape bulb of new T C securely to control end of old T C 10 From the top reach into ...

Page 7: ...point Press together To reset latched alarm or tune fail Press together briefly To run or Hold a program Press together for 3 seconds Notes If in difficulty by becoming lost in program mode press and together for 3 seconds to return to display mode check the INSTRUMENT ADJUSTMENTS above and try again When in program mode after 60 seconds of key inactivity the display will revert to either inPt non...

Page 8: ...arameter if calibrating unit value may vary ChEk off rEAd Var tECh CtA Ver 953 rSET none Level 2 SP1 P 0 hAnd off PL 1 100 PL 2 100 SP2 A Dv hi SP2 b none di SP 0 1 hi SC 100 Lo SC 100 in Pt tc t unit C Level 1 tune off band 0 3 int t off dEr t off dac 0 5 cyc t on of ofSt 0 0 SP Lk off SPrr 0 SPrn off SoAk Set 2 6 alarm deviation bnd 2 0 1 Cyc 2 on of ...

Page 9: ...low enough vacuum to remove moisture from the system it is recommended that the system be triple evacuated breaking each time with dry refrigerant Use care to purge the air from the system CHARGING REFRIGERATION SYSTEM Since capillary tube systems have small critical refrigerant charges we recommend that a field charge either be weighed in or put in from a portable charge cylinder After maximum va...

Page 10: ...g side 2 Low voltage to unit of protector 3 Overload protector defective 2 Determine reason and correct 4 Run capacitor defective 3 Check current replace protector 5 Excessive discharge pressure 4 Determine reason and replace 6 Compressor too hot return gas hot 5 Check ventilation restrictions in cooling medium restrictions in refrigeration 6 Check refrigerant charge fix leak if necessary Run capa...

Page 11: ... or by fire or flood CONSEQUENTIAL DAMAGES IN NO EVENT SHALL SCIENTEMP CORP BE LIABLE FOR SPECIAL INDIRECT OR CONSEQUENTIAL DAMAGES INCLUDING WITHOUT LIMITATION PRODUCT LOSS OR PRODUCT SPOILAGE CLAIMS NOR FOR ANY DELAY IN THE PERFORMANCE OF THIS WARRANTY DUE TO CAUSES BEYOND ITS CONTROL GENERAL The standard warranty and any service contract related to the STANDARD WARRANTY shall apply only to the ...

Page 12: ...I 2 I o r 1 L RED 2 115 VOLT BLACK gLARMLIGHT VOLT DC BLACK RC 34 TEMP S WITCH HTlEX BLACK l 1 ItpUT 1 IJHITE 415 617 8 RED I 2 3 6 VOLT BATTERIES RED T WHITE BLACK __ I RED J RED ALARM TEST S WITCH l RED RED BLACK ALARM ONIOFF SIJITCH POIJER RELAY 115 VOLT WHITE I T COI I I I FAN l III us VO I 0 WHITE CHART RECORDER 115 VOLT w HITE BLACK HIGH STAGE COMPRESSOR 115 VOLT LOIJ STAGE COMPRESSOR 115 VO...

Reviews: