5. Instructions for Use

 (cont.)

5.3 HYDR

IM

 M2 G4 Cycle Description Chart

The P3 and P4 disinfection temperature reflects the minimum guaranteed temperature 
at any time during disinfection anywhere in the wash chamber. However, the disinfection 
temperature is set to be around 93°C, with an upper switch point of the heater at 94°C 
and a lower switch point of 92°C.

The HYDRIM M2 G4 incorporates a unique design where the air from the dryer motor 
is pushed through the wash arms. The wash arms turn during the second high-speed 
drying phase. This unique feature allows the HYDRIM M2 G4 to eliminate the risk of 
water drops falling onto the dried instruments during unloading.

The disinfection time of P4 can be increased by 1 minute incremental intervals from 1 to 
5 minutes. The minimum disinfection time is 1 minute with a minimum A

0

 value of 3500.  

The minimum A

0

 value for disinfection time of 5 minutes is 6000. The final rinse time has 

to be set and validated by a technician after installation in the office.

NOTE:  ISO 15883-1 describes the time-temperature relationships for moist heat disinfection as follows: 

“For a moist heat disinfection process, a particular time at a particular temperature can be expected to have 
a predictable lethal effect against a standardized population of organisms. Definition of moist heat disinfection 
process may be achieved by means of the A

0

 method, which uses knowledge of the lethality of the particular 

process at different temperatures to assess the overall lethality of the cycle and express this as the equivalent 
exposure time to a specified temperature.”

* This is not a wash cycle. Always run a wash cycle following the rinse & hold cycle.
** Cycle times depend on the temperature and pressure of incoming water.
*** Water consumption includes drying phase

Cycle

Prewash

Wash

Rinse

Dry

Total Time

**

 

w/o Drying

Water

Consumption

P0 – Machine 
Cleaning Cycle

No initial draining.

<30ºC (cold)

3-10 minutes

(default 3 minutes)

N/A

<30ºC (cold)

2 minutes

N/A

6

minutes

16 L

P1 – Rinse and 
Hold Cycle*

Use to prevent soil from 
drying on instruments 
when they will not be 
washed within one hour.

<30ºC (cold)

3-10 minutes

(default 3 minutes

N/A

60ºC

1 minute

N/A

18

minutes

16 L

P2 – Regular Cycle

(no disinfection) – Use for 
moderately soiled loose 
instruments.

<30ºC (cold)

3-10 minutes

(default 3 minutes

50ºC

5-15 minutes

(default 9 minutes)

60ºC

1-10 minutes

(default 1 minute)

20-60 

minutes

(default 30 minutes)

35

minutes

32L

P3 – Heavy Duty 
Cycle

with disinfection

 – 

Use for heavily soiled 
instruments  
and cassettes.

<30ºC (cold)

3-10 minutes

(default 3 minutes

50ºC

5-15 minutes

(default 9 minutes)

90ºC

5 minutes

20-60 

minutes

(default 30 minutes)

60

minutes

40L***

P4 – Custom

(A

0

  between A

0

 = 3,500 

(rinse)  
and A

0

 = 6,000)

<30ºC (cold)

3-10 minutes

(default 3 minutes

50ºC

5-15 minutes

(default 9 minutes)

90°C

adjust. from  

1 to 5 

minutes

20-60 

minutes

(default 30 minutes)

40L***

Summary of Contents for HYDRIM M2 G4

Page 1: ...HYDRIM M2 G4 Operator s Manual Mode d emploi Bedienerhandbuch Manual del operador Manuale d uso HYDRIM M2 G4 Operator s Manual 95 113643 Rev 1 0 Copyright 2013 SciCan Ltd All rights reserved...

Page 2: ...e HYDRIM M2 G4 35 7 5 Filter and Wash Arm Maintenance 36 7 6 Annual Maintenance 36 8 Unit Testing and validation 37 9 Troubleshooting 39 10 Warranty 41 11 Specifications 42 12 Serial Number Plate 42 T...

Page 3: ...nd replacement instructions must be followed for the product to perform as designed All trademarks referred to in this manual are the property of their respective owners Contents of this manual are su...

Page 4: ...rface and or Hot Steam Caution Risk of electrical shock Disconnect power supply Caution Potential hazard to the operator The following symbols appear in this book This situation may lead to a mechanic...

Page 5: ...es and mouth Never sit stand or lean on the open door The unit may tip forward causing injury Always turn the unit OFF before adding softener salt adding solutions or performing routine maintenance to...

Page 6: ...rtaken before starting the machine See section 6 for care and maintenance details A separate pre installation checklist should have been supplied by your dealer Please review this prior to approving i...

Page 7: ...ethylene engineered foam and wood pallets Some municipalities operate packaging collection points with dedicated disposal bins to accommodate various materials What to do with your old unit A decommis...

Page 8: ...M M2 G4 such as the chamber show any of these spots or stains a water test may be required to determine the cause The installation of a water treatment system may be required to reduce the amount of d...

Page 9: ...top maximum 1 kg 2 2 lbs with salt 8 Close by screwing the lid tightly back into place and run a complete cycle with no instruments An improper seal can lead to corrosion Please note The water test s...

Page 10: ...ps scissors curettes and other stainless steel instruments free of deep cavities For best results instruments processed in the HYDRIM M2 G4 should have the following properties Heat resistance to a te...

Page 11: ...PN 01 110409S Closed hinged instruments cannot be cleaned Place hollow lumens suction tubes and other hollow instruments in the rack for vertical instruments PN 01 01 110411S Do not place hollow lumen...

Page 12: ...sinfection process may be achieved by means of the A0 method which uses knowledge of the lethality of the particular process at different temperatures to assess the overall lethality of the cycle and...

Page 13: ...commended loading of general instruments and the worst case loading for which the unit has been type tested and validated Handle all instruments with care to prevent personal injuries through puncture...

Page 14: ...pluggers curettes excavators etc For example Bone files pluggers curettes excavators etc For example dental handpieces turbines hollow lumens etc Handpiece holder washes or thermally disinfects the ou...

Page 15: ...allation technician may be required and is recommended If one of these factory validated loads is used by the dental office and this can be documented by pictures and or appropriate lists an initial o...

Page 16: ...ts Quarter size 1 4 Part number 01 109967S Rack for four 11 x 14 280 x 368 mm cassettes for upper or lower trolley Full size 1 1 Part number 01 109965S Statim 5000 basket Part number 01 107241 Hinged...

Page 17: ...HYDRIM M2 G4 Touchscreen overview Settings Unlock door Network connectivity green when active USB status green when active Programs Water softener detergent level and door status 5 Instructions for Us...

Page 18: ...creen Instructions Instructions Delay Remote Access Prepare for Shipping Set Printer Language Country Date Time Unit No Network Setup Reset Drying Counter Reset LCS Counter Screen Saver Temperature C...

Page 19: ...tup menu 5 9 Setting the time 1 Date Time 2 Scroll to Date Time and select TIME 3 From the TIME screen use the keypad to set the time Press EN to save and to return to the Setup menu NOTE If the HYDRI...

Page 20: ...ng the keypad select a maximum of 3 digits to be used as the unit s identifier number Press EN to save and to return to the Setup menu 5 12 Resetting the drying counter The drying counter must be rese...

Page 21: ...these steps NOTE Turning OFF the button sound does NOT turn off other alarms and cycle notification beeps 1 2 Scroll to Beep ON OFF and select 3 Use to scroll through your ON or OFF options and select...

Page 22: ...centers Should you need to adjust the contrast for your office follow these steps 1 2 Scroll to LCD Contrast and select 3 Use to scroll through your contrast options When you have found the contrast...

Page 23: ...screen you can assign up to four PINs To assign a PIN follow these steps 1 2 Scroll to User and select 3 To assign a user PIN select User PIN and use the numeric keypad to enter a number up to 4 digi...

Page 24: ...located at the back of the unit To connect your HYDRIM to a network using a router follow these steps 1 Connect your network cable to the Ethernet port at the back of the unit If your office uses a ro...

Page 25: ...Connection time will vary depending on your network speed and making an initial connection can take longer 5 24 Connecting to a wireless network The HYDRIM can be configured for wireless use by connec...

Page 26: ...jected cycle completion 8 The cycle is complete When the cycle is complete the program name and Cycle Complete is displayed and the unit beeps three times Press OK to open the door Use gloves to unloa...

Page 27: ...menu functions To access the user or setup menus follow these steps Chamber cleaning This cycle is used to periodically remove hardwater deposits from the chamber walls and racks Pour 1 0 litres of vi...

Page 28: ...ot Instruments that have not completed the cycle should not be used and should be reprocessed When replacing the kickplate ensure that the grounding wire is attached Grounding wire Ring 5 Instructions...

Page 29: ...2 The unit will record up to the last five successful cycles and the last five incomplete cycles If you select a cycle from the list it will display cycle information in a format similar to how it wou...

Page 30: ...rom the web portal 3 When the activity light on your USB storage device stops blinking or the USB icon on the LCD turns from a flashing green to a solid grey remove the USB storage device and transfer...

Page 31: ...l Printer if connecting a printer or USB Flash MSD if connecting the SciCan Data Logger Press the back arrow to return to the Setup Menu 4 In the Setup Menu select Baud rate 5 Select the rate required...

Page 32: ...undamaged gloves made of Natural or butyl rubber nitrile or neoprene Surgical gloves are ideal Always wear eye protection Protect exposed skin on arms If solution is spilled on clothing remove the ite...

Page 33: ...cleaning solution follow these steps 7 2 1 Replacing the HIP Cleaning Solution The HYDRIM M2 G4 uses a 3 8L 1 US gallon container of HIP Cleaning Solution Part number CS HIPL Only use the recommended...

Page 34: ...tion container by rotating forwards into the correct location in the solution door see picture 6 11 Close the cleaning solution door 12 Press OK on the touch screen The machine display will indicate P...

Page 35: ...log the unit The water softener salt is also available from SciCan Please contact your local dealer 7 3 Changing the Air Filter When the message Replace Air Filter appears the HEPA filter is in need o...

Page 36: ...ce call was performed Filter Maintenance Inspect the filter in the bottom of the chamber daily for debris and clean if necessary To clean remove the filter rinse under a tap and replace Ensure that th...

Page 37: ...icians Failure to maintain this equipment may invalidate the results of the following testing regime The recommended periodic testing protocols are as follows Commissioning Installation testing and va...

Page 38: ...eck for debris See section 6 5 Door seal check Visually check door seal for debris or wear If seal is worn or damaged call your dealer for technical service Load carrier check Visually check load carr...

Page 39: ...ater is not hot enough Check that the incoming hot water temperature is in the recommended range Check that the hot and cold hoses are not reversed No detergent message comes on when there is still ch...

Page 40: ...vice CF13 Temperature validation error Second temperature sensor readings are out of limits Call for service CF14 Water temperature too high Water in prewash stage too hot Check water connections at b...

Page 41: ...invoice identifies the item by serial number and clearly states the date of purchase No other validation is acceptable After one year all SciCan s warranties and other duties with respect to the quali...

Page 42: ...0 Hz see serial number label for requirements specific to your unit Rated load 2 5 kW Circuit breaker 13 amps Operating temperature range 5 C to 40 C Maximum relative humidity 80 for temp up to 31 C 5...

Reviews: