SciCan HYDR
IM
®
L110W G4
Quick Reference Guide
Operator’s Manual should be consulted before using this washer.
Close the door.
Select cycle and
press START.
Power unit ON.
Load Instruments.
Rinse and Hold Cycle
Prevents soil from drying on instruments.
Incomplete loads that will not be washed
within the hour.
NOT a cleaning cycle.
Regular Cycle
Moderately soiled loose instruments.
Heavy Duty Cycle
Heavily soiled instruments and cassettes.
CF2 Chamber Filling
Water is not filling the chamber
Check that the water hoses are not kinked.
Check that water shut-off valves are opened.
Call for Service.
CF15 Water Reservoir Overflow
Overflow switch triggered
Restart the program.
Call for Service.
CF21 Dosing System
Detergent is not dispensing
Tighten detergent cap.
Check for leaks.
Call for Service.
CF28
No Water Pressure
Check the water shut-off valves are open.
Check water hoses are not kinked.
Call for Service.
BASIC OPERATION
TOUCHSCREEN
Replacing the HIP
™
Cleaning Solution
Recommended precautionary measures with regard to safe handling of HIP solution:
• Wear undamaged surgical gloves and eye protection.
• After handling the product, remove and dispose of the gloves and wash hands prior to removing eye protection.
Cleaning the Chamber
Refilling the Water Softener Salt Reservoir
TROUBLESHOOTING
4
1
1-800-667-7733
1-800-572-1211
www.scican.com
MSDS available for download at www.scican.com
HYDR
IM
is a registered trademark and HIP is a trademark of SciCan Ltd. Manufactured by SciCan Ltd. 1440 Don Mills Road, Toronto, Ontario, Canada, M3B 3P9.
P1
Rinse&Hold
P2
Regular
P3
Heavy Duty
USB
Door
Lock
User/Technical
Menu
Network
Connection
Water
Softener/
Detergent
Level/Door
Status
Enter
Back
2
Pour 1.0L of vinegar or citric acid-based instrument neutralizer into
the chamber and follow these steps:
1.
Unscrew lid.
2.
Fill to the top (max. 1 kg / 2.2 lbs.)
3.
Screw lid on tightly.
4.
Run a complete cycle with no instruments.
IMPORTANT!
If the salt container is being filled for the first time, it MUST be filled with
at least 1 litre of water before adding the salt.
Only tablet salt should be used in the softening unit. Fine grained salt
such as table salt should NOT be used as it could clog the unit.
The water softener salt is also available from SciCan.
Please contact your local dealer.
P0 Cleaning
IMPORTANT! Do a Chamber Cleaning cycle every 25 cycles.
Failure to clean the chamber can promote corrosion.
Refer to Operator’s Manual for a more comprehensive list. Call for service if suggested recommendations do not resolve the error message.
FOR DETAILED INFORMATION PLEASE SEE PROVIDED OPERATOR’S MANUAL
3
95-115870 NA ML R2
Spare Parts and Accessories
CS-HIPL
HIP Cleaning Solution (2 x 3.8L)
01-112594S
Water Softener Salt
01-108305S
Water Test Kits
WTI-SK
Wash Test Indicator - Starter Kit
WTI-RK
Wash Test Indicator - Refill Kit
Terminal sterilization of instruments
MUST
follow processing in the HYDR
IM
L110W G4.