SciCan HYDRIM L110w G4 Quick Reference Manual Download Page 1

SciCan HYDR

IM

®

 L110W G4

 

Quick Reference Guide

 

Operator’s Manual should be consulted before using this washer.

Close the door.

Select cycle and 
press START.

Power unit ON.

Load Instruments.

Rinse and Hold Cycle

Prevents soil from drying on instruments.
Incomplete loads that will not be washed 
within the hour.

NOT a cleaning cycle.

Regular Cycle

Moderately soiled loose instruments.

Heavy Duty Cycle

Heavily soiled instruments and cassettes.

CF2 Chamber Filling 

Water is not filling the chamber

Check that the water hoses are not kinked.

Check that water shut-off valves are opened.

Call for Service.

CF15 Water Reservoir Overflow

Overflow switch triggered

Restart the program.

Call for Service.

CF21 Dosing System

Detergent is not dispensing

Tighten detergent cap.

Check for leaks.

Call for Service.

CF28 

No Water Pressure 

Check the water shut-off valves are open.

Check water hoses are not kinked.

Call for Service.

BASIC OPERATION

TOUCHSCREEN

Replacing the HIP

 Cleaning Solution

Recommended precautionary measures with regard to safe handling of HIP solution:

•  Wear undamaged surgical gloves and eye protection.
•  After handling the product, remove and dispose of the gloves and wash hands prior to removing eye protection. 

Cleaning the Chamber

Refilling the Water Softener Salt Reservoir

TROUBLESHOOTING

4

1

1-800-667-7733

1-800-572-1211

www.scican.com

MSDS available for download at www.scican.com

 

HYDR

IM

 is a registered trademark and HIP is a trademark of SciCan Ltd. Manufactured by SciCan Ltd. 1440 Don Mills Road, Toronto, Ontario, Canada, M3B 3P9.

P1

Rinse&Hold

P2

Regular

P3

Heavy Duty

USB

Door 
Lock

User/Technical 
Menu

Network 
Connection

Water 
Softener/
Detergent 
Level/Door
Status

Enter

Back

2

Pour 1.0L of vinegar or citric acid-based instrument neutralizer into 
the chamber and follow these steps:

1. 

Unscrew lid.

2.  

Fill to the top (max. 1 kg / 2.2 lbs.)

3.  

Screw lid on tightly.

4.  

Run a complete cycle with no instruments.

IMPORTANT!

If the salt container is being filled for the first time, it MUST be filled with 
at least 1 litre of water before adding the salt. 

Only tablet salt should be used in the softening unit. Fine grained salt 
such as table salt should NOT be used as it could clog the unit. 

The water softener salt is also available from SciCan.  
Please contact your local dealer.

P0 Cleaning

IMPORTANT! Do a Chamber Cleaning cycle every 25 cycles. 

Failure to clean the chamber can promote corrosion.

Refer to Operator’s Manual for a more comprehensive list. Call for service if suggested recommendations do not resolve the error message.

FOR DETAILED INFORMATION PLEASE SEE PROVIDED OPERATOR’S MANUAL

3

95-115870 NA ML R2

Spare Parts and Accessories

CS-HIPL

HIP Cleaning Solution (2 x 3.8L)

01-112594S

Water Softener Salt

01-108305S

Water Test Kits

WTI-SK

Wash Test Indicator - Starter Kit

WTI-RK

Wash Test Indicator - Refill Kit

Terminal sterilization of instruments 

MUST

  

follow processing in the HYDR

IM

 L110W G4.