PEM Hydrogen generator quick instruction manual
Optional security H2 overflow consumption control
3 port
valve
H2 Flow
Control
Neutral
gas
(N2)
GC or other
hydrogen
consumer
Out put
Hydrogen
(H2)
How does it work:
There is an event output that is controlled by SET Flow, when real flow exceed
the set flow more than 60 second then system realize an overflow
(it may be of
broken column in GC oven or any other leak in the stream of hydrogen)
and then
the system stop generating hydrogen and activates a 3 port valve and purge the
oven or any other chamber with neutral gas such as nitrogen (N2) and issue
flowing error in error line 4 of LCD and sounds the alarm buzzer.
Over flow Error