Description, Data Sheet FCKL 1528 Page -2
t
Frequency range 9 kHz - 30 MHz
t
10 Hz - Frequency steps
t
Conducted interference measurement
with L.I.S.N.
t
Field strength measurement with
adapter.
t
Integrated power attenuator for
receiver protection.
t
Optional high level tracking generator
is ideal to measure
Lamp attenuation
acc. to EN 55015.
Also for filter attenuation, free area
attenuation and to drive power
amplifiers.
t
Manual Operation, semi-automatic
operation with xy-recorder and PC-
control via IEEE-bus using the
Schwarzbeck software.
t
Fast 100% CISPR Quasipeak, CAV and
CRMS measurement with VARISCAN.
For many decades, most of the
interference measuring receivers were
used in laboratories. They were operated
manually using their front panel. This type
of operation including front panel control
will still be there in the future, but PC-
control gives value added measurement
because of increased speed and better
documentation. The unique r. f. and
analogue circuits of the FCKL 1528 give
precise measurement with or without PC-
control. The receiver comes complete for
EMI-measurement, but can be equipped
with useful options.
Characteristics of the FCKL 1528
Unique R.F. - circuitry
u
Attenuator with r.f.-relays uses resistive
Π
-attenuators with 1 dB steps. Total
resistive attenuation is 95 dB.
u
Switchable 10 dB high-power-attenua-
tor with 10 W for safe measurement
with L.I.S.N.s up to 4 x 400 A.
u
5 Input filters. Shape factors optimised
for EMI - measurement.
u
CISPR standard filters with 200 Hz and
9 kHz / - 6 dB. Filters are classic
double tuned band filters.
u
Integrated 25 Hz / 100 Hz Pulse-
standard for CISPR Band A and B
similar to IGLK 2914 for calibration.
Error is compensated by a EPROM list.
u
Integrated (optional) tracking generator
with 120 dB
µ
V (1 V) / 50
Ω
for mea-
surement of
Lamp attenuation,
filter
attenuation, field attenuation with an-
tennas and amplifier drive.
High precision measurement
u
Meter with 2 large scales.
Linear voltage scale with 1 dB-scaling
for the amplitude range
-10 dB / 0 dB centre of meter +6 dB
according to EN 55014 C.2.1.
plus Logarithmic overview
-25 dB / 0 dB centre of meter +25 dB
u
12 Bit A/D-converter
Easy to use
u
Functional areas of controls and
displays.
u
Small size, moderate weight
u
u
Rugged Aluminium cabinet
u
u
Low heat dissipation
u
Due to effective shielding no problems
even when used in the shielding room.
Data Interface
IEC-Bus-Interface: Connector 24 sockets
Sub D-Connector 25 sockets
§
Supply Voltages d.c. +12 V / -12 V for
auxiliary equipment
§
XY-recorder control Frequency Ampli-
tude, Penlift
§
Output voltage of active Demodulator
(Envelope) for auxiliary or monitoring
with Oscilloscope
§§
Sub-D-connector 9 sockets for
L.I.S.N-control
BNC-Outputs
§
I. F.-Output
§
Tracking generator output 120 dB
µ
V
50
Ω
(optional)