Pressing the right key increases the time by one hour (max. 23). Pressing the right key
decreases the time by one hour (min. 0).
Press the key to save all parameters and to leave the date/time mode again. Date and
time are continuously updated.
The TENStem eco basic switches OFF automatically after two minutes if no key is pressed.
In this case none of the adjusted parameters are saved.
Memory query mode
First set the date and time (see above). The memory query
mode is used to read out saved therapy data from patients. By
pressing and holding the left and
E
key for about 3 seconds
you enter the memory query mode. If no values are saved, the
display shows “null” for about 2 seconds and the device returns
to standby mode automatically. If you have saved therapy data,
the display shows the parameters of the last therapy session as
shown below:
»
Used program = P01
»
Frequency = 100 Hz
»
Pulse width = 200 µs
»
Therapy duration = 0 min 44 s
»
Day of therapy = 5 (fifth day)
»
Therapy session no. = 3
You can save up to 90 therapy sessions respectively the first 3 of each day. With the left
key you can return to the values of the previous day (until you reach the first recorded
day). The left key lets you proceed to the next day (until you reach the current day).
The right key brings you to the parameters of the last therapy session of the respective
day (until you reach the first session of the respective day). The right key shows you the
parameters of the next therapy session (until you reach the last session of the respective
day).
Resetting the memory
Pressing and holding the
E
button for about 5 seconds will delete all saved data. The
display shows “Clr” for about 2 seconds and the device can be used for stimulation again.
The TENStem eco basic switches OFF automatically after two minutes if no key is pressed.