9.2 Guarantee
We give a guarantee of 2 years from the date of purchase on the XTR2. This guarantee does not
cover accessories such as cables, electrodes, battery, charger, etc.
9.3
Maintenance and cleaning
No special cleaning or care agents are required for the XTR2. Clean the stimulator with a soft,
fluff-free cloth. Please ensure that no moisture permeates the stimulator. If moisture does
permeate the stimulator, a technical check must be carried out before re-use.
9.4
Classification
In accordance with the Law on Medical Devices, the XTR2 is classified as being a Class IIa
medical device.
9.5 Technical
check
We recommend a technical check on the XTR2 every 24 months. This includes:
1. Confirming that user instructions have been included in the accompanying documentation.
2. Checking the equipment for completeness.
3. Visual check:
- for mechanical damage
- for damage to all cables and sockets
4. Functional Safety
- Checking the output signals with a load resistance of 1 k
Ω
real current and voltage
- Checking the frequency
- Checking the pulse width
These technical checks may only be performed by suitably qualified persons. The results must be
documented with the date and name of the person carrying out the check.
10 Delivery Content
Art.-No. Article
Quantity
104064 SPORÉCUP
XTR2
1
106351
Cable Type 5.15
2
283400
Self-adhesive electrodes 50x50 mm (4 pieces)
1
101062
Charger WK 112 for XTR2
1
Transportation
bag
1
Operating
instructions
1
SPORÉCUP XTR2 | Instruction Manual
11
11 Accessories
+