background image

KL 1600 LED – Operating instructions

5/10

The KL 1600 LED emits high-intensity visible light. Because light-absorbing materials have the 
physical property of converting incident light into heat, damage can arise to heat-sensitive or 
flammable light-absorbing materials. To avoid such thermal damage and the potential danger of fire 
or burns, please obey the following instructions:

Never cover up the light guide socket or light guide exit (danger of fire)! 

Never cover up the open light guide socket or light guide exit with your hands or other parts of 
the body (danger of burns)! 

When illuminating heat-sensitive or flammable light-absorbing objects (e.g. in microscopy), 
special care must be taken to ensure that an appropriate suitable distance between light guide 
and object, and a suitable lamp brightness level are selected so that no thermal damage occurs.

When the light source is switched on, all light guide exists not being used in the working 
procedure must always be a safe distance – at least 10 cm – away from heat-sensitive or
flammable light-absorbing materials (prevention of possible danger of fire). Care must, therefore, 
be taken that each light guide exit is at the above-mentioned safe distance away from, for 
example, dark/colored textiles and dark/colored wooden or plastic surfaces.

To avoid unnecessary stressing of biological tissue by illuminating with visible light, reduce the 
brightness and duration of illumination to the absolute minimum required level.

Please make sure that your KL 1600 LED cold light source is operated at the voltage 
stated on the model plate (10).

The light source has been developed for operation in dry rooms only! (see Point 6 “Technical 
data”)!

The instrument must not be used in explosive areas.

Please ensure the air vents (7, 9) are always kept free. In the case of insufficient cooling an 
integrated thermo switch continuously regulates the light intensity (see Point 4 “Troubleshooting)

Safe disconnection from the power supply occurs only by pulling out the mains plug.

The instrument may neither be opened nor dismantled. Technical modifications are forbidden. 
Repairs must only be carried out by the manufacturer or by its authorized customer service 
agency

Please ensure that every user of the system has quick access to these operating instructions.

The manufacturer is not liable for damage caused by failure to obey these instructions. 

Summary of Contents for KL 1600 LED

Page 1: ...LED Operating instructions Status April 2012 Distribution in the UK Ireland Lambda Photometrics Limited Lambda House Batford Mill Harpenden Herts AL5 5BZ United Kingdom E info lambdaphoto co uk W www...

Page 2: ...tents 1 Important information 4 2 Operation 6 2 1 Light guide connection 6 2 2 Power connection 6 2 3 Start up procedure 6 2 4 Light intensity setting 7 2 5 Filter slide 7 3 Maintenance 8 4 Troublesho...

Page 3: ...n off switch 2 3 3 Locking screw 2 1 4 Light guide socket 2 1 5 Carrying handle 6 Mains connection socket back of instrument 2 2 7 Ventilation grid back of instrument 1 8 Filter slide 2 5 9 Air vents...

Page 4: ...f Risk Class 2 The KL 1600 LED fiber optic light source conforms to the following European directives 2006 95 EC Low Voltage Directive 2004 108 EC EMC Directive 2002 95 EC RoHS The technical documenta...

Page 5: ...le danger of fire Care must therefore be taken that each light guide exit is at the above mentioned safe distance away from for example dark colored textiles and dark colored wooden or plastic surface...

Page 6: ...two guide slots 2 2 Power connection Insert the hollow plug of the supplied power cord into the plug in socket 6 of the light source Please take care to ensure that you only operate your KL 1600 LED l...

Page 7: ...an insert filter from the filter set available as accessory When operating the light source the filter slide must always be engaged in one of the end positions or in the rest position This is the onl...

Page 8: ...oltage Light intensity dropping Electronics overheated Switch off instrument Ensure adequate cooling switch on again after prolonged cooling time 5 KL 1600 LED accessories A wide range of accessories...

Page 9: ...perature decreasing linearly to 75 Air pressure hPa 700 1060 Transport and storage Temperature C Rel air humidity Air pressure hPa 20 70 10 95 non condensing 500 1200 Contamination level 2 Electrical...

Page 10: ...our SCHOTT product was produced and developed with high quality materials and components The symbol indicates that electrical and electronic devices must be separated from domestic waste and appropria...