OSSIBLE MALFUNCTIONS
Problem
Cause
Remedy
No function
• Power plug not plugged in
properly
• Fuse of the power supply has
tripped
• Controller not plugged in
properly
• Pull out the plug and insert it
correctly
• Check fuse, test unit on a differ-
ent elec. socket
• Note connection plan
No display function
• Controller cable disconnected
• Controller cable not connected
correctly
• Check the connection
• Note connection plan
No cooling
Compressor turned off
Temperature controller faulty
Temperature above 40 °C
Check supply and exhaust air
Contact the seller
Allow device to cool down to
below 40 °C
No heating
Power supply interrupted
Heating element is faulty
Temperature controller faulty
• Check power supply
• Contact the seller
• Contact the seller
Bad cooling
• Evaporator dirty
• Fan dirty
• Overheating on installation site
• Cleaning by service engineer
• Cleaning by service engineer
• Check supply and exhaust air
Display shows
"FIH" or "Fill"
• Broken sensor
• Contact the seller
Temperature varies for
more than 20° C
• Broken sensor
• Contact the seller
• Don’t further use the device to
prevent subsequent damage
The problems listed above are for guidance and must be seen as examples. Should these or similar prob-
lems occur, immediately turn off the device and discontinue use. Contact a qualified specialist or the
manufacturer immediately.
Noise during operation (only for compact system)
To keep the selected temperature constant when cooling, from time to time, the hot or cold combiplate
switches the compressor of the cooling unit on. The sounds which occur are related to functionality. They
will reduce automatically as soon as the unit has reached operating temperature. The humming noise
comes from the compressor.