background image

Tension Meter 

 
 

MKM Series 

 

 

Model   MKM-50

  

MKM-100

  

MKM-400

Edition 

MKM 03.3.E

Instruction Manual

Valid as of: 01.03.2007

 

• Please keep the manual for future reference!

S C H M I D T

c o n t r o l i n s t r u m e n t s

Summary of Contents for MKM-100

Page 1: ...sion Meter MKM Series Model MKM 50 MKM 100 MKM 400 Edition MKM 03 3 E Instruction Manual Valid as of 01 03 2007 Please keep the manual for future reference S C H M I D T c o n t r o l i n s t r u m e n t s ...

Page 2: ...4 1 8 Copyright 4 1 9 Declaration of Conformity 4 2 Available Models 5 2 1 Specifications 5 2 2 Delivery Incudes 5 2 3 Unpacking 5 3 Initial Setup and Operating Procedure 6 3 1 Operating Elements 6 3 2 Charging the Battery 7 3 3 Notes Before Starting Measurement 7 3 4 Operating Procedure 7 3 5 Verification of Calibration 8 4 Service and Maintenance 9 5 Cleaning 9 6 Verification Interval 9 6 1 Veri...

Page 3: ...the device Insufficient inspection of device components that are subject to wear Opening the device or improper repair work Disasters caused by the effects of foreign objects or by force majeure 1 1 Notices within the Operating Instructions The fundamental prerequisite for the safe handling of this device and its troublefree opera tion is the knowledge of the basic safety notices and safety instru...

Page 4: ...sion meter Any other use or any use exceeding this intention will be regarded as misuse Under no circumstances shall HANS SCHMIDT Co GmbH be held liable for damage resulting from misuse The intended use also includes Complying with all notices included in the Operating Instructions and observing all inspection and maintenance works 1 5 Training of the Personnel Only trained and instructed personne...

Page 5: ...0 01 N Model Tension Ranges cN Take up Speed m min SCHMIDT Calibration Material MKM 50 5 50 15 PA 0 12 mm Ø MKM 100 10 100 15 PA 0 12 mm Ø MKM 400 50 400 8 PA 0 20 mm Ø 2 1 Specifications Calibration According to SCHMIDT factory procedure Accuracy 1 full scale FS or 1 graduation on scale Take Up Speed 15 m min MKM 50 MKM 100 8 m min MKM 400 Scale Diameter 41 mm Temperature Range 10 45 C Air Humidi...

Page 6: ...edure 3 1 Operating Elements 1 HANDLE 8 Jack socket 6 35 mm ø for handle 2 MOTOR SWITCH 9 SCALE 3 Jack plug 6 35 mm ø 10 Tension meter 4 TAKE UP WHEEL 11 PROCESS MATERIAL 5 Charger connector 12 GUIDE EYE 6 Jack plug 3 5 mm ø 13 OUTER ROLLERS 7 Battery charger 14 MEASURING ROLLER ...

Page 7: ...ding on the orientation of use plug in the JACK PLUG of the HANDLE at either of the two JACK SOCKETS provided on the sides of the TENSION METER Thread the PROCESS MATERIAL through the GUIDE EYE the OUTER ROLLERS and the MEASURING ROLLER Then wind the PROCESS MATERIAL 2 or 3 times around the TAKE UP WHEEL Press the MOTOR SWITCH to start measuring The SCALE now shows the measured tension values 3 3 ...

Page 8: ...ase supply a material sample of 5 m for this purpose WEIGHT OUTER ROLLERS MEASURING ROLLER PROCESS MATERIAL SCALE fig 3 5 Verification Procedure Hang a WEIGHT which corresponds to the tension to be measured pay attention to the correct unit of measure from the PROCESS MATERIAL vertically as shown in fig 3 5 Always use a fresh portion of the material to be measured Before verifying the calibration ...

Page 9: ... OR LIABILITY shall be accepted for damage resulting from improper cleaning The question of finding the right frequency of calibration accuracy verification depends on several different factors Operating time and load of the SCHMIDT tension meter Tolerance band defined by the customer Changes to the tolerance band compared to previous calibrations Therefore the interval between verifications of ca...

Page 10: ...y customer No Yes No Yes No Yes Vertification of the calibration Visual inspection and functional test Values within specification Prepare certification for incoming calibration Determination of costs Make of cost estimate Repair and adjustment Verification of the calibration Calibration label on the instrument No Invoice scrapping and disposal Invoice repair and calibration services and dispatch ...

Page 11: ...ription to our service department Please indicate in your order whether you require an Inspection Certificate 3 1 according to DIN EN 10204 8 Repairs HANS SCHMIDT Co GmbH Schichtstr 16 84478 Waldkraiburg Germany Service address Shipping instructions We kindly ask for return free of charge for us if possible by airmail parcel All occurring charges if any such as freight customs clearance duty etc w...

Page 12: ...Thickness Gauge Yarn Package Durometer and Shore A Durometer Sample Cutter Balance Moisture Meter Leak Tester Softometer Mailing address P O B 1154 84464 Waldkraiburg Germany Shipping address Schichtstr 16 84478 Waldkraiburg Germany Hans Schmidt Co GmbH Phone int 49 0 8638 9410 0 Fax int 49 0 8638 4825 int 49 0 8638 67898 More than 70 years Worldwide e mail info hans schmidt com Internet http www ...