Schlumbohm EndoPilot2 Quick Start Manual Download Page 4

EndoPilot²

 

 

- 4 - 

ReFlex Calibration 

 

For the use of the ReFlex movement (Procodile files) an additional calibration is necessary, 
if your device was manufactured in 2018 (see sticker on the back of the control unit). 
 
The prerequisite is that a software update has been carried out beforehand and that the 
software version of your device is at least: v4.196. 
 
Calibration with dongle: 
 
1. Switch off your device 
 
2. Insert the dongle into the blue socket on the right side (motor connection) 
 
3. Switch on your device - the calibration is done automatically 
 
4. Check

:  

 
• 

Leave the dongle inserted. 

 

• 

Select the "Service" menu (gear symbol at the top left 

of the start screen)

 

• 

Select "Service Information" below.

 

• 

In the Motor line you will find the following values on the 

far right: RF_L: 60 (+/-2) and RF_R:60 (+/-2) Note: If the 
dongle is removed, other values may be displayed 
 