Lubrication
Lubricate the dry medical device immediately after cleaning and/or disinfection.
Recommended lubrication cycles:
•
Essential after every internal cleaning
•
Before each sterilization or
•
After 30 minutes of use or at least once daily
•
Chucking system once a week
1
.
Lubricate the contra-angle with oil, for
example with W&H Service Oil F1, MD-
400. Please follow the instructions of use
printed on
the
aerosol can and on the
packaging.
Or
2.
Lubricate the contra-angle in an automatic
instrument maintenance device, for
example in the W&H Assistina 301plus (see
instructions for use coming with the device).
After maintenance, please place the contra-angle in an upright position on a clean
disposable cloth. This way, excessive oil may drip off so that it cannot get into the
motor.
Testing after Lubrication
1. Hold the contra-angle with the head pointing
downwards.
2. Let the contra-angle run briefly so that excess
oil can escape. To do so, please start the
calibration function of the EndoPilot. In case of
dirt leaking from the contra-angle, the entire
hygiene and maintenance procedure must be
repeated.
3. Wipe the contra-angle with disinfectant wipes.