8
Calibration
8.1
Test equipment for the calibration procedure
Drawing 4
8.2
Proof of the quality of the ESD-pulse
The shape of the output current has to be proofed with an oscilloscope with a bandwidth of at least
2 GHz. The test bed has to be set up as shown in the drawing above. For further information’s concerning
the normative probe please have a closer look into the standard IEC 61000-4-2 / EN 61000-4-2 and the
ESD verification set.
The most important parameters like rise time and pulse width are defined with a tolerance of ± 25% and
± 30%.
Therefore we recommend a calibration every two years !!
9
Test setup
The correct test setup can be found in the standard:
a)
Industrial
IEC 61000-4-2
EN 61000-4-2
b)
Automotive
ISO 10605