![Schiller MAGLIFE LIGHT User Manual Download Page 34](http://html1.mh-extra.com/html/schiller/maglife-light/maglife-light_user-manual_1212088034.webp)
MAGLIFE light
4.7.2 Device configuration menu
The configuration menus are hidden during normal operation. Access is
initiated by pressing the navigation button
(6)
when the device is switched
on and keeping it pressed until the configuration menu is displayed.
Note:
To exit the
Device configuration
mode, the device must be
switched off.
4.7.2.1 Alarm limits submenu
The default thresholds are set for the three type of patients - Adult, Child
and Neonate. Select one type of patient, press button
(6)
and make the
setting in the following tab
Select the parameter by navigation button
(6)
counter clockwise and
confirm by pressing button
(6)
. The first value will flash. Turn as required to
modify the value. Save the value by pressing button
(6)
. The next value is
automatically selected. When the last parameter value has been entered,
the selection cursor goes back to the parameter selection column.