Cleaning and Disinfection
14
User Guide
Cleaning Interval
14.2
A
rt.
no.
:
2.
511
335
Rev.
a
CARDIOVIT CS-104
14.2 Cleaning Interval
The recording device comes into contact with the patient, and it is recommended that
it is cleaned after each use. Any visible soiling must be cleaned immediately.
14.3 Cleaning / Disinfecting
Observe the following safety notes when cleaning/disinfecting the device.
–Never immerse the device or patient cable in liquid
–Never pour or spray liquid directly onto the unit patient cable
–Make sure that no liquid penetrates the connections or openings
–The device and / or patient cable must not be autoclaved or sterilised with steam
Before cleaning/disinfecting the unit thoroughly inspect for any signs of damage and
any improper mechanical function of buttons or connectors. Switch off the recorder
before cleaning.
Do not spray the device directly.
Make sure that no liquid penetrates the device.
Clean the recorder with a damp cloth
slightly moistened (not wet)
on the
surface only. Use cleaning agents that are mild and diluted with water that are
suitable for PC polycarbonate - approved cleaning solutions are listed following,
Use a clean lint-free cloth moistened with detergent and wipe the unit to clean. Leave
to dry in the air for at least 30 minutes
Ensure liquid does not get into connectors. If liquid should get into connectors, dry the
area with warm air, and then check the equipment to confirm that it operates properly.
Wipe off any excess cleaning solution. Do not let the cleaning solution run into or
accumulate in connector openings, latches, or crevices. If liquid penetrates the device
or into the connectors this may interfere with correct functioning. Remove the patient
cable, the memory card and the battery. Leave the recorder in a warm, dry room with
the battery chamber open for 48 hours and then check the equipment to confirm that
it operates properly. If the functioning is still affected, contact the manufacturer.
Ensure the contacts of the patient cable are completely dry before reassembling