Troubleshooting
ARGUS VCM
Error Codes
When
ARGUS VCM
detects an error condition, the monitor will attempt to show an error
code on the display screen.
If such an error occurs during the monitoring operation, the monitor will sound a low-
priority alarm. Audible alarm can be terminated if the alarm silence button pressed, but
it depends on error codes and conditions.
Table 10 provides a complete list of error codes and problem identification.
Table 10. Technical Error Codes
Error Codes
Description
E01 Battery
failure
E02
Abnormally shutdown last time
E09
System error (malfunction)
E21 Air
leakage
E22
Cuff not detected
E23
Kinked or neonate hose
E24
Cuff too large for neonate
E25 Overpressure
condition
E26 Motion
artifact
E27
Weak pulse or no pulse detected
E28
Valid BP not found
E29
NIBP module error (malfunction)
E41 Sensor
off
E42 Sensor
disconnected
E43 Bad
sensor
E44
Loss of pulse
E49 SpO
2
module error (malfunction)
“P”
Loss of tissue contact
E62 Probe
disconnected
E63 Probe
error
E64
Probe heated too high
E65
Ambient temperature out of range
E69
Temperature module error (malfunction)
42
ARGUS VCM
Service Manual