Service Mode and Demo Mode
Introduction
Service Mode
Demo Mode
Introduction
This section describes Service mode that allows authorized personnel to review/check
system version, battery voltage level, NIBP inflation cycle, total system runtime, factory
default reset, night panel, NIBP calibration, NIBP pressure test and sound calibration in
order to obtain service-related information about the monitor. Also this section explains
how to demonstrate the monitor (Demo mode).
Service Mode
Follow the procedure below to enter Service mode.
1. With the monitor powered off, press Power button and NIBP start/stop button
simultaneously.
2. System version displays on the monitor.
3. Press Mode button to select a mode item.
System Version
The revision level of the system software displays: system software version, NIBP
module version, SpO
2
module version and Temperature module version.
Figure 5. System Version
ARGUS VCM
Service Manual 25
SpO
2
module version
System version
Temperature module version
NIBP module version