Further hygiene measures
Fig. 4
Fig. 5
Cover the central section and
turn the vaporizer on. The
water will have evaporated
after around 10 – 15 minutes
and the disinfection process
will be concluded.
In the home:
After sealing the socket with
the stopper the hand unit
and individual parts can be
cleaned in a dishwasher
(open chamber side facing
downwards).
In clinics:
After sealing the socket with
the stopper the hand unit
can be disinfected / sterilized
through immersion disinfec-
tion, placing it in a chemical-
thermal disinfector (60 °C for
10 minutes) or in a gas sterilizer.
The nebuliser head, valve,
mouthpiece and baffle plate
can be disinfected, gas-steril-
ized or autoclaved at up to
134 °C as described above.
After cleaning and disinfection of the unit in a vaporizer / dishwasher or after chemical-thermal disinfec-
tion the unit must be allowed to cool for approximately 30 minutes before removing the parts and start-
ing the next inhalation procedure.