Portable analyzer
6 of 18
GA-S1
Version 2015
4.1 Checking the analyzer
All portable analyzers are being calibrated before delivery.
To carry out and obtain reliable measurements the analyzer has to be checked regularly. This can be
done by comparing the values of the analyzer with a calibrating gas, see chapter 5: Control procedure.
If the given values on the display
do not
match with the values of the calibrating gas, calibration is
necessary, see chapter 6: Calibration.
In the event of no sample gas suction, the suction hoses and the connected filter have to be checked.
The filter could be blocked, in which case it has to be replaced.
The user is entirely responsible if the user carries out calibration himself.
In this case, Schele Electronic cannot be held to account for any damages.