Measuring procedure
1. Switch on the device by pressing the
power button ( )
for 3 sec.
2. Select the right calibration curve using
the buttons
or
. The name of the
calibration curve is shown at the top of
the display. Please see the instructions
for selection of calibration curve below.
3. Hold the device in one hand and press it
onto the paper roll (at the breadth) resp.
onto
the
paper
pile
with
gentle
compacting pressure (1,0 kg).
For
measurement at running rolls the PM4
(contact-free temperature measurement) is
necessary!
4.
Now the display shows the water
content.
The temperature is shown on
the left.
5. To save the results in the save menu
press the
( button). The storage
was successful when the number in
front of the symbol
increased. To
reach the store menu please press the
left ( ) button.
6. The menu level „Options -> Log time”
offers the possibility to activate the menu
item „3 sec.“. If you press the
(
button) in the measuring window after
activation, the device will save a
measuring value every 3 seconds.
7. Pressing the button
( button)
completes the saving series. Now you
can name the saved values using the
arrow keys (supplier 1).