100

%

Mn

Mt

Mn

F

 

Calibration curves 

Calibration curves 

Declaration 

Measuring  range 

Wood chips 

Standard wood chips  

10 - 50 % 

Coarse chips 

Coarse wood chips 

10 - 50 %

 

Industrial chips 

Industrial wood chips 

10 - 50 %

 

Test block 

! Only for testing the BLL with the test block !

 

 

 

Wood  chips:

  standard  chips  of  wood  (forest  wood  chips) 

according to standard EN ISO 17225-1 class 

P16, P31

 and 

P45

.  

 

Coarse  chips:

  for  coarse  wood  chips 

P45  or  P31  but

 

with 

fewer fines. 

 

Industrial chips: 

for industrial chips of wood 

without barks and 

fines

 (similar 

P45

 or 

P63

)       

 

 

Definition of wood chips classes (according EN ISO 17225-1) 

The  stated  numbers  refer  to  the  particle  size  that  goes  through 
round gaps of the corresponding diametres (e.g. P16: 16 mm).  

o

 

P16

 

minimum 75% of the bulk is between 3.15 and 16 mm 

o

 

P31 

minimum 75% of the bulk is between 8 and 31.5 mm 

o

 

P45 

minimum 75% of the bulk is between 8 and 45 mm 

o

 

P63 

minimum 75% of the bulk is between 8 and 63 mm 

 

Compression of wood chips  

The humimeter BLL is calibrated for normally compressed wood chips. If the 
measured  wood  chips  are  compressed  to  a  much  lesser  or  greater  extent, 
this  will  cause  measuring  imprecisions.  Normally  compressed  wood  chips 
are defined in norm EN 15103 (determination of the bulk density). 

 

Determination of the material reference moisture 

The  principle  is  a  comparison  measurement  with  the  dehydration 
method  according  to  EN  14774.  Take  the  measured  sample  and 
weigh it. Dry it out in an oven and weigh it again.  

 

   

M

n

:   Mass with average moisture content 

M

t

 :   Mass of the dried sample  

%F:  Calculated absolute moisture  

Summary of Contents for humimeter BLL

Page 1: ...User manual Moisture meter with insertion probe for determination of water content of wood chips humimeter BLL Version 2 4_en Schaller GmbH 2015...

Page 2: ...page 2 Version 2 4_en Design of the device Measuring head Water content Batteries slot flipside Temperature Insertion probe Grap bar...

Page 3: ...r key The name of the calibration curve can be seen at the head of the display 4 Plug the probe of your humimeter BLL straight into the wood chips It is not allowed to load the measuring head incorrec...

Page 4: ...of the corresponding diametres e g P16 16 mm o P16 minimum 75 of the bulk is between 3 15 and 16 mm o P31 minimum 75 of the bulk is between 8 and 31 5 mm o P45 minimum 75 of the bulk is between 8 and...

Page 5: ...r Keypad symbols Measuring window Rolling Menu Power ON OFF Switch upper Switch lower Save Hold Watch the saved data Suppliers data can be added Rotate display Menu Enter Switch upper Switch lower Exi...

Page 6: ...w batteries in the device Make sure that the position of the battery poles is correct 5 Press down the batteries and close the cap If the battery symbol appears in the measuring window resp if a criti...

Page 7: ...ars Press the key The measuring value remains on the display until another button is pressed Display lighting Press the key Backlight will turn off automatically after 30 seconds Backlight will be act...

Page 8: ...ing on the device Your humimeter automatically starts the main menu Activate the the super user functions according to the instructions above Navigate to Options o Userlevel and confirm by pressing th...

Page 9: ...ng the data transfer on the humimeter Press the key until you reach the menu see image on the right Then choose Send Logs and confirm by pressing the key Now choose Manual Logs and confirm with again...

Page 10: ...BLL Then switch on the device Not till then the other side of the cable has to be plugged in at the printer Switch on the printer by pressing Now the green LED is blinking If it does not blink please...

Page 11: ...otection IP 40 Scope of supply 4 x 1 5Volt AA Alkaline Batteries Rubber protective housing grab bar Options Wooden case Test block USB data interface modul Thermo printer runs by battery Exemption fro...

Page 12: ...ation curve Before you measure your sample double check the correct selection of the calibration curve Wet or mouldy material Frozen measuring material Water film at the measuring head After measuring...

Reviews: