humimeter FS1 & FS1.1 Operating Manual
Page 16
5.
The measuring process
5.1
Preparing a measurement
To do so:
The device has to have nearly the same temperature than the product being
measured. It is recommended to let your humimeter device adjust to the surrounding
temperature of the material being measured for at least 30 minutes.
1. Place the empty and clean measuring cup (0.25 liter)
on the switched-off scale (figure
). Then switch on the
scale.
»
The scale must display 0.0 g with the empty measuring
cup on it. The measuring cup must not be weighed.
2. Check whether the measuring chamber of the device is
empty. When the device is switched on, there must not
be any material in it.
»
Empty the instrument and clean the measur-
ing chamber if necessary (see
).
).
4. Perform the automatic calibration (see
).
5. Select the required product type (see
) by pressing the
or
button (see
).
9
10
11