Fiber Collimators Series 60FC-SF
Adjustment_60FC-SF 09.12.2020 © Schäfter+Kirchhoff GmbH
Schäfter+Kirchhoff GmbH Kieler Str. 212 22525 Hamburg, Germany +49 40 85 39 97-0 [email protected] www.sukhamburg.com
1
Fiber Collimators Series 60FC-SF
This document provides assistance in installing the fiber collimators series 60FC-SF. It describes how a fiber
cable is attached and how the collimation setting is adjusted.
1. Before You Start
The fiber collimator is shipped with a protection cap for the fiber receptacle and with a front cap.
Please remove all the protection caps first and do not use them as beam
dumps (risk of photo contamination).
Notice:
There is a two-part protection cap on the receptacle. For removing this cap perform the following steps in
oderto remove this cap:
Figure 1:
First, remove the threaded cap from
the fiber receptacle of the collimator.
Figure 2:
Then, slightly loosen the small pin
screw. Use the screwdriver 9D-12.
Make sure to not loosen it too far, as it
is small and easily lost.
Figure 3:
Now please pull out the plug.