![Schaffner NSG 438 User Manual Download Page 50](http://html1.mh-extra.com/html/schaffner/nsg-438/nsg-438_user-manual_1209940050.webp)
NSG 438
44
10.4.
NSG 438 system error messages
Nr.
Text
Explanation
Action
006
INTERLOCK
OPEN
The "Interlock-circuit"
is open.
Press interlock button, or close the
interlock circuit at the back of the
base unit.
115
EUT FAILURE
The connected EUT
has signaled a fault.
EUT input has detected an EUT
fault. Reset EUT first and then
press return on screen.
125
HV SUPPLY
TIMEOUT
The HV cannot be
loaded in the specified
time.
HV voltage module has detected
during measurement an
uncertainty. Switch off the base
station wait for 10s, switch on
again.
126 HV HOLD
TIMEOUT
Unit stops after 30s
without discharge.
Press return on screen. Restart
the test.
127
HW FAULT
An internal fault on the
processor board has
been detected.
Switch off the base station wait for
10s, switch on again and continue
testing. If error persists, contact
your nearest Schaffner service
centre.
129
A FIELD IS
SELECTED
The "RUN" or the "HV-
ON" key has been
pressed although an
operator field is still
selected.
Finish the input in the operator
field and then start with “RUN”.
201
HV INTERNAL
DISCHARGE
During test, an internal
discharge has been
detected.
Press return on screen and trigger
again. If error persists, contact
your nearest Schaffner service
centre.
202
BATTERY
EMPTY
The battery is low and
needs to be charged.
Recharge battery with the original
power supply.
210
ERROR ADJUST
FAILURE < xxx >
HV module can not
reach the max. voltage
during calibration
procedure.
The voltage can be selected to the
displayed maximum value.
Contact you nearest Schaffner
service centre.
211
FAULT
DETECTED
The software has
found erroneous
behaviour during
generation of the
pulse.
Stop the test. Switch off the base
station wait for 10s, switch on
again and continue testing. If error
persists, contact your nearest
Schaffner service centre.
217
VOLTAGE TOO
HIGH
The selected value is
too high.
Reduce the voltage level.
247
HV TRAFO TOO
HOT
NTC resistor too hot
after endurance runs.
Power off the NSG 438 and wait
about 1h.