51
MPS4264
Section 5: Software
Validation
DESCRIPTION
Starts a validation and places the MPS into VAL mode
SYNTAX
VAL [<F or FM>]
Where:
F- Initiates a single temperature field validation, with no oven connection. Connects to the
calibrator.
FM- Initiates a single temperature manual field validation with no oven or calibrator connec-
tion.
When no parameter is present, the unit performs an automated validation based on the cali-
bration group settings
RETURNS
Validation Data
EXAMPLE
VAL
>
Multiple Unit Validation
DESCRIPTION
When the master receives the MVAL command it starts the multiple device validation sequence
described in the section on multiple device calibration and validation.
SYNTAX
MVAL [<F or FM>]
Where:
F- Initiates a single temperature multi-unit field validation, with no oven connection. Connects
to the calibrator
FM- Initiates a single temperature mult-unit manual field validation with no oven or calibrator
connection.
When no parameter is present, the unit performs an automatid multi-unit validation based on
the calibration group settings.
RETURNS
Validation Data
EXAMPLE
MVAL
>
Zero Offset Calibration
DESCRIPTION
Starts a zero calibration and places the MPS4000 into CALZ mode.
SYNTAX
CALZ
RETURNS
A prompt when ready for the next command.
EXAMPLE
CALZ
>
Summary of Contents for MPS4264
Page 1: ...MPS4264 Miniature Pressure Scanner Hardware and Software Manual Software Version 2 07...
Page 2: ......
Page 3: ......
Page 6: ...INTENTIONALLY LEFT BLANK...
Page 12: ...6 MPS4264 Section 1 Specifications INTENTIONALLY LEFT BLANK...
Page 18: ...12 MPS4264 Section 2 Introduction INTENTIONALLY LEFT BLANK...
Page 32: ...26 MPS4264 Section 4 Hardware INTENTIONALLY LEFT BLANK...
Page 38: ...32 MPS4264 Section 5 Software INTENTIONALLY LEFT BLANK...
Page 81: ...75 MPS4264 Section 6 Maintenance INTENTIONALLY LEFT BLANK...
Page 86: ...80 MPS4264 Section 6 Maintenance INTENTIONALLY LEFT BLANK...
Page 87: ...81 MPS4264 Appendix INTENTIONALLY LEFT BLANK...