AccuPro-5000
™
MEASUREMENT SYSTEM OPERATOR MANUAL
Issued by Department:
Engineering
Doc. Number:
5000-00-00-002
Revision:
02
Effective Date:
10/05/2021
File: 5000-00-00-002-r2_
AccuPro_5000_Operator_Manual
Copyright© 2021 Scaletron Industries, Ltd
All Rights Reserved
Page
49
of
63
Figure 31: TARE SECURED Control Flow Diagram
a.
Access to this menu requires that channel calibration procedures be completed. User is prompted
via the display to first enter any missing required parameters.
b.
If the system is in Run Mode with Tare Secure enabled (see
will appear as “Tare Secured”, otherwise, it will read “Tare Adjust.”
c.
For a dual-channel system, first select the desired channel for which to conduct the procedure. The
number of any chosen active channel will be indicated in the upper-left corner of the display for the
remainder of this submenu.
d.
If the system is in RUN Mode with Secure Tare enabled (See
be able to make adjustments to the current tare value. Instead, the display will indicate the current
tare value.
e.
Select and enter the desired tare units. Units of Pound (lb), Kilogram (kg), Gallon (gal), and Liter
(ltr) are supported. If a unit change is detected, the default tare value is reset to zero. Otherwise,
the starting tare value is set to the previously entered value. Note that the selected tare units do not
have to match the selected scale measurement units.
f.
Enter a Tare value in the previously-selected units using one of the two methods previously
described. The maximum Tare value is 0 (no Tare) and the minimum value is the negative
equivalent of the previously entered Capacity Weight converted into the selected units of tare. An
arrow symbol is appended at the end of the top row of the display to indicate the available range of
the tare adjustment. For example, a Down-Arrow indicates the tare value is currently at its
maximum, or zero, meaning only a decrease in value is possible. Similarly, an Up-Arrow indicates
the current value is at its minimum, meaning only an increase in value is possible. An Up/Down-
Arrow indicates the current value can be increased or decreased.
8.10
Relay Setpoint
Select this menu option by pressing the Enter Key while the display appears as shown in Figure 32.
Control Flow if Secure Tare is
“ENABLED”
Previously
Entered Value in
Selected Unit
Path taken when ENTER Key pressed.
Path taken when ARROW Key pressed.
EXIT
1: TARE SECURED
0.0 lb