18
Setup of units of
measurement to
be displayed
1
81.S.u.
|
85.S.u.
☆
3 20 [Pcs]
☆
4 1F [%]
☆
5 00 Unit not set
Display of the
auxiliary scale
interval
9.Ai
0 Disable
☆
1
Enable (
is the auxiliary scale interval.)
Note: The auxiliary-scale-interval place does not
represent a verification scale interval. Use it
only as a reference value.
GLP-compliant
print
0.GLP
☆
0 Disable Prints a GLP-compliant format at calibration.
(Refer to
“9.3 GLP-compliant print” (Page 31).)
1 Enable
Output format
while the auxiliary
scale interval is
displayed
A.PrF.
1
No output is made while the auxiliary scale interval is
displayed.
2
Output is made even while the auxiliary scale interval is
displayed.
☆
3
Output is made with
“/” added to the left of the
auxiliary-scale-interval place.
☆
1
~☆
5: default factory settings [81.S.u.]-[85.S.u.]
※
1
Can be set only for a model on which the lock switch is off.
For a model on which the lock switch is on, only the units set when the switch was off are effective.
4.3 Interface Section
Displayed when [7.
I.F.
] is set to [1] or [2]
Item
Set Value
Description
Output Control
71.o.c.
0 Stop output
1 Output continuous at all times
2 Output continuous if stable (stop output if unstable)
3
Outputs once by pressing Memory key (irrespective
of whether stable).
4
Outputs once if stable. Outputs if the balance is stable
when a sample is loaded after the preceding sample
has been removed and the balance indicated zero, or
less.
5
Outputs once if stable, and stops output when unstable.
Even if the sample is not replaced, the balance is output
once when it stabilizes next time (including the zero
indication).
6
Outputs once if stable, and outputs continuously when
unstable. Even if the sample is not replaced, output of
the balance stops when it stabilizes after being output
once.
☆
7
Pressing Memory key causes the balance to output
once when stable.
Baud Rate
72. b.L.
☆
1 1200 bps
2 2400 bps
3 4800 bps
4 9600 bps
Parity
73. PA.
☆
0 None
Displayed only when [7. I.F. 2
(7-digit numeric format)] is specified.
1 Odd
2 Even
☆
denotes a factory-setting
※
The data interval in continuous output mode is 0.1 to 1 second. (The interval varies depending
on weighting conditions and other factors.)