9
Reference to standards
During the certification process, a fixed number of
instruments was subjected to stringent clinical tests. These
instruments used the same program for measuring the
blood pressure as the instrument which is now in your
possession. This instrument is therefore deemed to have
been clinically tested as well.
The instrument has been produced in accordance with the
European standard for blood pressure manometers: EN
60601-1 and EN 1060-1 thru EN 1060-3.
Electromagnetic compatibility:
The instrument satisfies the requirements of the European
standard EN 60601-1-2.
The requirements of EU directive 93/42/EEC for medical
products of class IIa are also met.
Warranty
Our blood pressure monitor has two years’ guarantee. A
defective instrument will be repaired free of charge during
the warranty period. After the warranty period has expired
you can also have a defective instrument repaired by our
Customer Service, for which you will get an estimate in
advance free of charge.
The warranty does not cover any damage that has been
caused by improper use, accidents, wear and tear (cuff),
non-compliance with the instructions for use or if the
instrument has been tampered with. The packaging, the
batteries and these instructions for use are excluded from
the warranty.
12
A colored LCD display will assess the measured blood
pressure in three steps:
green light: ........... optimal blood pressure
yellow light: ......... normal blood pressure
red light: ............. borders on hypertonia or hypertonia
Assessment of blood pressure values
The WHO classifies high blood pressure in three degrees
of severity:
optimal blood pressure
below 120/80
normal blood pressure
below 130/85
bordering on hypertonia
below 140/90
1
st
degree: mild hypertonia
below 160/100
2
nd
degree: moderate hypertonia
below 180/110
3
rd
degree: severe hypertonia
above 180/110
Care and maintenance
♥
Avoid extreme temperatures, moisture, dust and the
direct exposure to sunlight.
♥
The cuff incorporates a sensitive, air-tight bubble.
Please avoid any stress by turning or bending it.
♥
Clean the instrument with a soft, dry cloth. Never use
petrol, thinner, or similar solvents. Stains on the cuff
can be gently removed with a wet cloth and soap.
Do not wash the cuff!
♥
Do not drop the instrument or treat it violently in any
other way.
♥
Never attempt to open the instrument, as the calibration
set by the manufacturer will become invalid.
10
Device Description
Description of LCD
1.
Date
2.
Time
3.
Systolic Pressure
4.
Diastolic Pressure
5.
Heartbeat Symbol
6.
Pulse Display
7.
Low Battery Symbol
8.
Memory Set
9.
Memory Symbol
10. Alarm Symbol
11. Average Measurement
Symbol
12. Inflation/Deflation
Symbol
5
Set the alarm clock:
1. Press the ”MODE Button” to enter into the alarm setting
mode.
2. Continue to press the ”MODE Button” to select the
alarm to adjust.
3. With the chosen alarm screen displayed,
the hour will begin to flash at the middle of
the screen, Fig 4. Press the ”MEM Button”
to change the hour. Once the correct
hour is displayed, press the ”SET Button”
to confirm.
4. After the hour is confirmed, the minutes
will begin to flash, Fig. 5. Press the
”MEM Button” to change the minutes.
Once the correct minutes are displayed,
press the ”SET Button” to confirm.
5. The alarm feature must
be turned on immediately
after setting the time for
the alarm. Press the
“MEM Button” to turn the
alarm on, Fig. 6. Press the
“MEM Button” a second time to turn the alarm off, Fig. 7.
Press the “SET Button” to confirm and turn the unit off.
7