Currency Discrimination Fitness Sorter SB – 2000EX Service Manual
57
MR SHADING NG
NORM : 0.09
NOISE :
1
LEVEL : 370 (HIGH) (1)
3
MR TEST
ERROR
MR SHADING OK
NORM : 1.07
NOISE :
4
LEVEL : 370(HIGH) (1)
44
MR SHADING
PLACE SHADING PAPER
ON HOPPER
When CIS cannot recognize the
shading sheet(1US$ bill) this error
happens. Make sure to check the
CIS lens, harness connection
condition, whether harness is cut
off or not
SHANDIG NG (Case 1)
Harness cut off or disconnected
between
CF AMP
and
Image Board
.
(No Signal: Normal level values are not
transferred to the Image Board)
Check if MR SHADING result is
“OK”
If it is “NG”, repeat placing the note
several times until “OK” shows up
on the display.