SBIG SG-4 Quick Start Manual Download Page 1

 

 

 

 
 
 
 
 

 

Application Note 

SG-4: Installing Jumpers for  

+12V and Ground on the Guide Port 

May 25, 2010 

 

Introduction 
The SG-4 provides mechanically isolated relays to connect the +X, +Y, -X, and –Y signals to the 
Common input on the Guide Port.  This is sufficient for most applications and all consumer telescope 
guide ports SBIG is familiar with.   
 
As shipped from the factory, the Guide Port pin-out is as shown here: 

 

 
When one of the guide signals (+X, +Y, -X, -Y) activates, the SG-4 connects the signal to the Common 
pin on the Guide Port.  The Common pin and the NC pin have no other connection to the internal 
electronics of the SG-4. 
 
However, some systems, such as the SBIG Relay Box, may require power (+12V) and/or Ground on the 
Guide Port.  The Common pin can be connected to the SG-4 Ground and the NC pin can be connected to 
the SG-4 +12V power signal if necessary. 
 
This application note describes how to install jumpers in the SG-4 to c12V and Ground to the 
Guide Port. 
 
NOTE: Be aware that this procedure is not recommended for most users and should only be done if you 
know for certain that you need +12V and/or Ground on the Guide Port.  If you are not sure, DO NOT 
follow this procedure. 
 