Text of the Instruction Manual in HELP Menu 

1   Main function of the KERN SSG Software for Gold Tester     

    

With this software, the individual ultra sound velocity of your gold piece can be calcu-
lated.  
 
Gold pieces as bars or coins are typically not made of pure gold with a sound velocity 
of 3240 meter per seconds (m/s). Often copper or other components are in small 
shares elements of gold pieces. This is necessary to obtain a better physical charac-
teristics and it needs to be respected in the check of genuineness. 
 
The components of the mixture of your gold piece can be taken off the expose that 
comes with the bar or the coin. Or it can be asked for at the manufacturer.  

2    Start the software 

 
Extract    

 

KERNSSG.zip 

 
Start 

 

 

KERNSSG.exe 

 

3   Start of the procedure  

 
Please select „File“ then „New“ 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Summary of Contents for TD GOLD 40

Page 1: ... D 72336 Balingen E Mail info sauter eu Tel 49 0 7433 9933 199 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Goldtester SAUTER TD GOLD 40 Version 1 3 05 2018 GB TD_GOLD_40 BA e 1813 PROFESSIONAL MEASURING ...

Page 2: ... for Gold Tester 3 2 Start the software 3 3 Start of the procedure 3 4 Choice of the outer appearance 4 5 Weight and Dimensions 4 6 Safety level 5 7 Material Composition 5 8 Takeover of sound velocity to the instrument 6 9 Measurement with the TD special gold 6 10 Interpretation of the measured value 6 11 Alternative measurement procedures 7 12 Instruction Manual of TD US as a base for using TD GO...

Page 3: ...ocity of 3240 meter per seconds m s Often copper or other components are in small shares elements of gold pieces This is necessary to obtain a better physical charac teristics and it needs to be respected in the check of genuineness The components of the mixture of your gold piece can be taken off the expose that comes with the bar or the coin Or it can be asked for at the manufacturer 2 Start the...

Page 4: ...of gold test object The following procedure for the measurements is recommended Determination of the weight Type approved balance e g KERN KB 650 2NM Turn the balance on Warm up for approx 30 minutes Place the test object on the weighing pan Read the result in grams Determination of the length dimensions Caliper Take the thickness dimension Read the result in millimetres ...

Page 5: ...el of 95 This can be overwritten by any other value 7 Material Composition Pure gold bars or coins are rather uncommon Typically gold pieces are a composi tion of gold and other materials e g to provide a higher stability That means that other materials are added in the production process Common are amongst others silver and copper The exact composition of the gold piece can be seen in the exposé ...

Page 6: ...eys and the sound velocity can the changed in 10 m s steps Please round the calculated sound velocity accordingly By pressing VEL again this setting is stored 9 Measurement with the TD special gold Please put a little bit coupling gel ATB US 03 can be reordered onto the test object Place the sensor on the gel gently pressed by hand The sensor has to be connect ed with the TD display unit If the co...

Page 7: ...of TD US as a base for using TD GOLD 40 12 1 Features The exclusive Micro computer LSI offers high measurement accuracy The instrument offers high power of emission and a wide spectrum of receiving sensitivity Sensors of different frequencies can be identified Rough surfaces even cast iron can be measured It is used in almost all kinds of industries Convenient to measure the thickness of many mate...

Page 8: ...ting conditions Temperature 0 to 50 C Humidity 80 Dimensions 120 x 62 x 30mm 4 7 x 2 4 x 1 2 inch Weight about 164g batteries not included Accessories Carrying case Instruction manual Ultrasonic sensor 12 3 Front panel description 3 1 Sensor plug 3 2 Display 3 3 mm inch key 3 4 Power key 3 5 Material selection key 3 6 Plus key 3 7 Ultrasonic sensor 3 8 Calibration key 3 9 Minus key 3 10 Battery co...

Page 9: ...Plus key 3 6 or the Minus key 3 9 has to be pressed to select the material code to measure Then the material selection key 3 5 has to be pressed to confirm The instrument changes into the measuring mode and on the display occurs 0 If a material code is selected without confirming this selection the instrument will automatically change back into the measuring mode after a few seconds In this case t...

Page 10: ...AL key 3 8 has to be pressed to confirm Then the instrument changes back to the measuring mode d Like this calibration has been finished and automatically saved in the instrument 12 6 Measurement procedure a The Power key 3 4 has to be pressed to switch on the instrument b The mm inch key 3 3 has to be pressed to select the correct measurement unit c The sensor 3 7 has to be placed onto the materi...

Page 11: ... the same as the known thickness In this case the set value is the velocity of the material to be measured With this any number of unknown thicknesses of the same material can now be measured e To change velocity VEL key 3 13 has to be pressed To return into the measuring mode this key has to be pressed again or it has to be waited until the instrument automatically shows 0 f By using velocity mea...