background image

 

 

TD_US-BA-e-2020 

6. 

Measuring procedure 

 

1.  The power button 3-4 is pressed to switch on the unit. 
2.  The mm/inch button 3-3 is pressed to select the correct measuring unit. 
3.  The sensor 3-7 is placed on the material surface to be measured, provided that 

the material code has been selected correctly. Make sure that the "Measuring 
process  in  progress"  indicator  is  visible  and  the  character 

((-))

  3-11  appears. 

The display shows the measurement result. 

4.  The measurement result is retained until a new measured value is measured. 

The last value remains on the display until the instrument is switched off. 

5.  The device can be switched off either by pressing the on/off key or by the auto 

power  off  function,  which  becomes  active  one  minute  after  the  last  key  is 
pressed. 

7. 

Measurement with adjustment of the ultrasonic speed 

 

1.  When the VEL- key 3-13 is pressed, the display shows the last stored ultrasonic 

speed. 

 

2.  Coating  thickness  &  material  thickness  measurement  at  known  ultrasonic 

speed:  
The speed of sound can be adjusted by pressing the plus or minus key. This 
changes the value shown on the display up or down.  The increase is initially 
10m/s. If the Plus or Minus key is pressed down for more than 4 seconds, the 
increase is 100m/sec each time. 

 

3.  A little contact gel is applied to the material to be measured. Now the sensor 3-

7  is  pressed  onto  the  surface  to  be  measured.  On  the  display  the  coating 
thickness can now be read, provided that the coupling symbol appears on the 
display.  If  the  ultrasonic  velocity  of  a  certain  material  is  known,  it  is  easy  to 
measure the coating thickness by using step 7b). 

 

 

4.  Layer  thickness  &  material  thickness  measurement  at  unknown  ultrasonic 

speed: 
Take  a  material  sample  with  known  layer  thickness  or  material  thickness. 
Repeat the 1st step (setting the ultrasonic speed) and 7c) until the measured 
value  corresponds  exactly  to  the  known  layer  thickness.  In  this  case,  the  set 
value  is  the  ultrasonic  velocity  of  the  material  to  be  measured.  Now  any 
unknown layer thickness of the same material can be measured. 

5.  To  change  the  ultrasonic  speed,  press  the  VEL  button  3-13.  To  enter  the 

measuring mode, press this key again or wait until the instrument automatically 
displays '0'.  

 

Summary of Contents for TD 225-0.1US.

Page 1: ...Balingen e mail info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Ultrasonic Material Thickness Gauge SAUTER TD US Version 2 0 04 2020 GB TD_US BA e 2020 PROFESSIONAL MEASURING ...

Page 2: ...ou will enjoy your quality measuring device with its wide range of functions For questions wishes or suggestions we are at your disposal Table of contents 1 Functions 3 2 Technical data 3 2 1 Scope of delivery 3 3 Description of the control panel 4 4 Material selection 4 5 Calibration Adjustment 5 6 Measuring procedure 6 7 Measuring with adjustment of the ultrasonic speed 6 8 Battery replacement 7...

Page 3: ...ing the material thickness of many materials e g steel cast iron aluminium copper brass zinc quartz glass polyethylene PVC grey cast iron nodular cast iron Automatic shutdown function to maintain power Connectable to a PC for data transmission with RS232C cable and software available as optional accessory 2 Technical data TD US Display 4 digits 10mm LCD display Measuring range 1 2 225mm 45 steel M...

Page 4: ... the material selected by the user The cdxx material relationship is as follows No Code Material 1 2 3 4 5 6 7 8 9 10 11 12 cd01 cd02 cd03 cd04 cd05 cd06 cd07 cd08 cd09 cd10 cd11 xxxx Steel Cast Iron Aluminium Copper Brass Zinc Quartz glass Polyethylene PVC Grey cast iron Ductile Cast Iron Sound velocity 3 1 Sensor plug 3 2 Display 3 3 mm inch button 3 4 On and Off Button 3 5 Material selection bu...

Page 5: ...cted and stored it is stored in the instrument memory As long as no change is made the meter always uses this material code 6 To enter the material code selection menu press the selection button 3 5 To leave the menu again press the selection button 3 5 again or wait until the device after a few seconds returns to the measuring mode and the display shows 0 5 Calibration Adjustment 1 A little conta...

Page 6: ... can be adjusted by pressing the plus or minus key This changes the value shown on the display up or down The increase is initially 10m s If the Plus or Minus key is pressed down for more than 4 seconds the increase is 100m sec each time 3 A little contact gel is applied to the material to be measured Now the sensor 3 7 is pressed onto the surface to be measured On the display the coating thicknes...

Page 7: ...y symbol appears on the display the batteries should be replaced 2 The battery cover is stripped off the meter and the batteries are removed 3 The batteries are inserted by observing the polarity when inserting them 4 If the device is not used for a longer period of time the batteries should be removed Note To view the CE declaration please click on the following link https www kern sohn com shop ...

Reviews: