TB-BA-e-1713
7
The black chequered shaft segment is seated movable on a spring. By means of the
spring, the sensor tip presses onto the test object with a defined force. This way,
measurement errors will be avoided.
It is recommended to perform several test measurements before the first use of the
instrument. This way, further measurement errors will be avoided.
10 Trouble-shooting
10.1 The instrument should always be calibrated on the same base material used for
the actual measurement, instead of the base plate included in delivery. By doing this,
measurement accuracy will be better right from the start.
10.2 The sensor will eventually be worn off. Its life will generally depend on the num-
ber of measurements taken and also on the surface roughness of the coating. Re-
placement of a sensor should only be done by trained and experienced staff.
10.3 The sensors of coating thickness gauges should only be replaced by identical
types or series. Otherwise, measurement accuracy may be affected or the gauge
could become damaged, in the way that it can no longer be used.
Annotation:
To have a look at the CE Declaration of Conformity, please click onto the following
link: