sauter CO Y5 Instruction Manual Download Page 3

CO Y5-BA-e-2120.docx 

 

Short description 

  Stainless steel IP66 (acc. to EN 60529) tension / compression load cell 

  Suitable for test stands, force gauges, automation plants, etc. 

  Accuracy OIML R60 G1 

  Very low height 

  Option on request accuracy C4 

  RoHS conform  

Available models 

Capacity 

Model 

0,5 kg (

≈5N) 

CO 0.5-Y5 

1 kg (

≈10N) 

CO 1-Y5 

5 kg (

≈50N) 

CO 5-Y5 

10 kg (

≈100N) 

CO 10-Y5 

 

Introduction 

Please pay attention to the notes of the operating manual: read these operating manual 
carefully  before  commissioning,  even  if  you  already  have  experience  with  SAUTER 
measuring cells. 
After  receiving  the  measuring  cell,  it  should  be  checked  in  advance  whether  no 
transport damage has occurred, whether the outer packaging or other parts or even 
the article itself has been damaged. If any damage is evident, please inform SAUTER 
GmbH immediately. 
 
 

Warning notes 

Observe the national accident prevention regulations. 
Not  correct  executed  use  of  the  measuring  cell  can  lead  to  serious  injuries,  death, 
material damage and personal injuries. Use only by trained and experienced people. 
Never load more than the max. capacity to a measuring cell. Overloaded measuring 
cells no longer have the required accuracy. Overloaded or deformed measuring cells 
are no longer allowed to use and must be changed quickly. 
Never step under suspended loads.  
Always fasten overload- or break security to your construction. 
Pay attention to the allowed static and dynamic loads of the used equipment. 
Let the measuring cells calibrate at regular intervals.  
Only load the measuring cell in its specified load direction. Avoid lateral forces.  
Control the measuring cells to deformations and cracks. 
Measuring  cells  with  4-Conductor  cable  will  change  their  characteristic  value  when 
shortened or extended the cable length.  
 

Summary of Contents for CO Y5

Page 1: ... GmbH Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Data sheet SAUTER CO Y5 V 1 0 06 2021 GB CO Y5 BA e 2120 docx ...

Page 2: ...Y5 V 1 0 06 2021 Instruction Manual Data sheet Summarize 1 Short description 3 2 Available models 3 3 Introduction 3 4 Warning notes 3 5 Warranty 4 6 Detailed specifications 4 7 Wiring 5 8 Dimensions in mm 5 9 Certificate of Compliance 5 ...

Page 3: ...t please inform SAUTER GmbH immediately 4 Warning notes Observe the national accident prevention regulations Not correct executed use of the measuring cell can lead to serious injuries death material damage and personal injuries Use only by trained and experienced people Never load more than the max capacity to a measuring cell Overloaded measuring cells no longer have the required accuracy Overlo...

Page 4: ...ns Output sensitivity FS mV V 1 0 10 2 0 10 Maximum capacity Emax N 5 10 50 100 kg 0 5 1 5 10 Combined Error FS 0 1 Temp effect on span F S 10 C 0 05 Temp effect on Zero F S 10 C 0 05 Minimum dead load Of Emax 0 Safe overload of Emax 200 Ultimate overload of Emax 500 Zero balance of FS 2 Excitation recommended voltage V 10 15 Input resistance Ω 680 10 Output resistance Ω 680 10 Insulation resistan...

Page 5: ... 4 conductor cable Shield not connected to element 4 wire diagram 8 Dimensions in mm 9 Certificate of Compliance To have a look at the CE Declaration of Conformity please click onto the following link https www kern sohn com shop de Downloads ...

Reviews: